11. Function check<br />
11.1 Checking that the nurse call system is free of<br />
faults<br />
11.1.1 Fault displays on equipment for the nursing staff<br />
74<br />
Faults are shown in the display (”Fault”, ”FLT” or ”FR”) of the following devices,<br />
along with the fault location:<br />
On corridor displays (scrolling) and alternately with calls and presences.<br />
On the terminal SD6 P (when presence is activated) when scrolling through<br />
the current messages or directly when there are no calls and presences.<br />
On the display module (when presence is activated), alternately with calls<br />
and presences.<br />
Here, the faults are displayed in the same way as calls, i.e. initially only in their<br />
own ward or, in the case of group coupling, in the coupled wards. If the fault is<br />
not remedied within the call overflow time, it is displayed in all wards in the build‐<br />
ing.<br />
11.1.1.1 ConLog PLUS Management Software<br />
All of the nurse call system's fault messages can be identified in the<br />
ConLog PLUS management software by looking at and evaluating the call docu‐<br />
mentation of the messages present.<br />
11.1.2 Fault messages on equipment for technicians<br />
11.1.2.1 Group Electronic<br />
Button 1<br />
Button 3<br />
Button 5<br />
Button 2<br />
Button 4<br />
Button 6<br />
Fault on group bus when LED in button 5 is permanently on or off.<br />
Fault on ward bus when LED in button 6 is permanently on or off. The output<br />
to the fault signal lamp also switches.<br />
Group Electronic faulty or is not completely programmed when LED in but‐<br />
ton 2 is on. The fault signal relay (potential-free) also switches.<br />
CONCENTO PLUS - <strong>Technical</strong> <strong>Manual</strong>, 11/12 (Rev. 3.0), 00 8814 14