You also want an ePaper? Increase the reach of your titles
YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.
5.2 Nurse call system with speech: Low call volume<br />
g (equipotential bonding)<br />
g (equipotential bonding)<br />
g (equipotential bonding)<br />
c (group bus) c (group bus)<br />
c (group bus)<br />
Reserve<br />
Reserve<br />
Reserve<br />
19 0800 84<br />
19 0700 10<br />
19 0700 10<br />
19 0700 10<br />
19 0700 00<br />
Functional ground<br />
Cables legend:<br />
c IY(ST)Y 4x2x0.8<br />
k NYM 2x2.5 mm 2<br />
or NYM 2x1.5 mm 2<br />
g NYM 1x1.5 mm 2<br />
77 3400 00<br />
230 V AC<br />
24 V<br />
DC<br />
77 3400 00<br />
230 V AC<br />
77 3400 00<br />
230 V AC<br />
CONCENTO PLUS - <strong>Technical</strong> <strong>Manual</strong>, 11/12 (Rev. 3.0), 00 8814 14<br />
c<br />
5. Installation examples<br />
19 0782 08<br />
19 0700 20 19 0700 20 19 0700 20<br />
Staff room<br />
19 0700 20 19 0700 20<br />
24 V<br />
DC<br />
c<br />
19 0782 08<br />
19 0700 20 19 0700 20 19 0700 20<br />
Staff room<br />
19 0700 20 19 0700 20<br />
24 V<br />
DC<br />
c<br />
analog a/b<br />
Paging/DECT<br />
ESPA 4.4.4<br />
19 0803 00<br />
LAN LAN for<br />
remote maintenance<br />
Order no. Product name<br />
19 0700 00 Management Interface<br />
19 0700 10 Group Electronic (GE)<br />
19 0700 20 Terminal SD6P<br />
k<br />
c (ward bus)<br />
k<br />
k<br />
c (ward bus)<br />
k<br />
k<br />
c (ward bus)<br />
k<br />
230 V AC<br />
19 0700 20 19 0700 20 19 0700 20<br />
19 0700 20 19 0700 20 19 0700 20<br />
19 0700 20 19 0700 20 19 0700 20<br />
19 0700 20<br />
Staff room<br />
19 0782 08<br />
19 0700 20 19 0700 20<br />
Note: The bus terminating resistor must be ac‐<br />
tivated in the first and last bus user of the group<br />
bus and ward bus.<br />
Order no. Product name<br />
19 0803 00 ConLogPLUS Management Software<br />
19 0800 84 Fire alarm interface<br />
19 0782 08 Corridor display, 8‐digit, double-sided<br />
77 3400 00 Power supply unit UPS<br />
39