Contura Z Helium Leak Detector<br />

Leak Detectors<br />

Technical Data<br />

Contura Z<br />

Smallest detectable leak rate<br />

Automatic test sequence, test cycle duration approx. 15 s mbar x l x s -1 < 10 -7<br />

Continuous testing, test cycle duration 40 s mbar x l x s -1 < 10 -8<br />

Leak localization with sniffer probe mbar x l x s -1 < 10 -7<br />

Test cycle duration<br />

Base setting s approx. 15<br />

Continuous measurement (adjustable) s 999<br />

Measurement range<br />

Qualitative measurement mbar x l x s -1 1.0 x 10 -8 - 9.9 x 10 -1<br />

Quantitative measurement mbar x l x s -1 > 9.9 x 10 -1<br />

Recorder outputs V lin: per decade 1-10 mantissa<br />

Staircase voltage for the exponents V log: 0-10 total meas. range<br />

Relay outputs Pcs. 3<br />

Ready:<br />

Measurement output ready<br />

Fail:<br />

Error message<br />

Meas:<br />

Trigger leak rate pass / fail<br />

Power consumption kVA 1.7<br />

Ramp-up time (after startup) min < 3<br />

Usable diameter of foil chamber<br />

Up to a product height of 20 mm mm 540<br />

Up to a product height of 100 mm mm 460<br />

Maximum product size (W x D)<br />

Up to a product height of 20 mm mm 380 x 380 / 250 x 470<br />

Up to a product height of 100 mm mm 325 x 325 / 250 x 380<br />

Power supply, single-phase<br />

230 V ± 5%, 50 Hz<br />

230 V ± 5%, 60 Hz<br />

Dimensions (W x H x D) mm 790 x 1210 x 1230<br />

Weight kg (lbs.) 166 (365.2)<br />

B5<br />

Ordering Information<br />

Contura Z<br />

Part Number<br />

Leak testing unit EURO 230 V / 50 Hz 122 60<br />

Leak testing unit 230 V / 60 Hz 122 61<br />

MONITOR software 122 65<br />

Calibrated test leak 122 67<br />

Test diaphragm (50 pcs.) for measuring helium concentrations 122 68<br />

B5.29