Settings and Measurement<br />

Switching displays<br />

Analog and Logic/Pulse can be switched as shown below.<br />

• Display Logic/Pulse Data<br />

A screen switches<br />

with ENTER key.<br />

• Display Analog Data<br />

Analog settings<br />

This screen allows you to set conditions for analog signals.<br />

CHECKPOINT<br />

ALL and Span All Settings are set only for the currently displayed group (per 10 channels).<br />

(1)-1 Input<br />

Selects input conditions.<br />

Selection<br />

Off<br />

Voltage<br />

Temperature<br />

Humidity<br />

Description<br />

Disables the measurement of input signals and the waveform and digital display.<br />

Used for measuring direct-current voltages.<br />

Used for measuring temperatures.<br />

Used for measuring humidity with humidity sensor B-530.<br />

This selection sets the voltage range to 1V and disables EU settings.<br />

(1)-2 Range<br />

Specifies the range of signal input to be measured.<br />

Input setting<br />

Selection<br />

Voltage 20, 50, 100, 200, 500mV; 1, 2, 5, 10, 20, 50, 100, 200, 500V; 1-5V<br />

Temperature TC-K, TC-J, TC-T, TC-R, TC-E, TC-B, TC-S, TC-N, TC-W,<br />

Humidity No selection<br />