Settings and Measurement<br />

(3)-4 Pre-trigger<br />

Specifies the percentage of data to be captured before a trigger is generated.<br />

* Pre-trigger can be selected only when the data capture destination is the internal RAM and the<br />

start side source setting is other than OFF.<br />

If the pre-trigger is 30%<br />

Time available for data capture<br />

30%<br />

Start side trigger<br />

(3)-5 Repeated Capturing<br />

Sets the repeat function for repeated capturing.<br />

Selection<br />

Off<br />

On<br />

Description<br />

The repeat function is not used.<br />

The repeat function is used.<br />

After one data capture process ends, the next data capture process starts (or a trigger is waited<br />

for if the start side source setting is other than Off).<br />

(3)-6 Repeat Interval<br />

Specifies the interval from the start of one data capture process to the start of another data capture process<br />

if the repeat function is used.<br />

(3)-7 Timer Trigger Information<br />

Displays the timer and trigger information in an easy-to-understand manner.<br />

Example<br />

Timer Mode : Every Day Cycle<br />

Start Side Source Setting : Level<br />

Stop Side Source Setting : Time<br />

Repeated Capturing : On<br />

A trigger generated in this interval<br />

will not start data capture.<br />

Data capture is enabled only in this interval.<br />

Data is actually captured in this interval.<br />

* This is only a conceptual illustration in which the graph length, number of times of repetition, and<br />

other items do not agree with the actual time.<br />

(3)-8 Alarm Setting<br />

Specifies alarm generation conditions and output destinations. When the conditions specified here<br />

are met, an alarm is output from the alarm output terminal (specify the output destination number<br />

for each channel).<br />

With "Hold Generated Alarm" set to "Hold", the alarm status is maintained after the conditions are<br />

met once even if they do not continue to be met (Press the CURSOR key to clear the alarm).<br />

See the next page for the condition setting for each channel.<br />