IMPROVED GCMS ANALYSIS OF THC IN SALIVA<br />

Charles R. Lyle, Matthew P. Clabaugh, F. Roark Galloway*: Shimadzu Scientific Instruments, Columbia,<br />

MD; Nadina Giorgi: Redwood Toxicology Laboratory, Santa Rosa, CA<br />

One <strong>of</strong>the most challenging GCMS applications is the determination <strong>of</strong>THC in saliva. Detection for THC<br />

at low ppb [nglml] levels is required due to several factors including; I) rapid elimination <strong>of</strong>THC in saliva<br />

from time <strong>of</strong> exposure, 2) low amount <strong>of</strong> drug for 'active' dose and 3) small sample volume available at<br />

time <strong>of</strong> collection. This application demonstrates an example <strong>of</strong> how the Shimadzu GCMS QP-2010 can<br />

provide the instrument performance necessary to complete this challenging analysis. Samples were<br />

collected using the Quantisal saliva collection kit. Extractions are performed by extraction with<br />

hexane:ethyl acetate [9:1] with 3 % acetic acid. Derivatization is completed using 50 uL BSTFA to form<br />

the trimethylsilane derivative. 1.0 uL <strong>of</strong> sample extract is injected for analysis. Using the standard SPL­<br />

2010 split/splitless injection port on the GC-2010 as a high pressure injector, the method was performed<br />

with a splitless injection using a Restek 3.4mm. id. SIL TEK liner. With high-pressure injection <strong>of</strong> 175 kPa,<br />

fast temperature ramping <strong>of</strong>35 C/min and high column flow rate, the peaks eluted from the column into the<br />

mass spectrometer ion source in less than 3.00 minutes. These results were obtained using a 15m X<br />

0.25mm X .25 urn phase DB-IMS column. Calibration from 1.0 25 ppb THC produced a linearity<br />

coefficient <strong>of</strong> 0.9999 by internal standard calculation using THC-D3 at 12.5 ppb. Preliminary LOQ is<br />

determined to be 0.25 ppb. Precision at 3.0 ppb THC is 5.97%.<br />

1.0 ppb THC [TMS] Standard Calibration, 1.0 - 25 ppb THC<br />

AnaR.tliu<br />

2.<br />

,.<br />

y2 = 0.99986<br />

Keywords: THC, Saliva, GCMS<br />