Chapter 5 Error MessagesExecution Errors-113 Undefined headerA command was received that is not valid for the function generator.You may have misspelled the command or it may not be a validcommand. If you are using the short form of the command, rememberthat it may contain up to four letters. Example: TRIGG:SOUR BUS-121 Invalid character in numberAn invalid character was found in the number specified for a parametervalue. Example: *ESE #B01010102-123 Exponent too largeA numeric parameter was found whose exponent was larger than32,000. Example: BMOD:NCYC 1E34000-124 Too many digitsA numeric parameter was found whose mantissa contained more than255 digits, excluding leading zeros.-128 Numeric data not allowedA numeric parameter was received but a character string was expected.Example: DISP:TEXT 123-131 Invalid suffixA suffix was incorrectly specified for a numeric parameter. You mayhave misspelled the suffix. Example: SWE:TIME 0.5 SECS-138 Suffix not allowedA suffix was received following a numeric parameter which does notaccept a suffix. Example: BM:PHAS 10 DEG (this command does notallow a suffix).-148 Character data not allowedA discrete parameter was received but a character string or a numericparameter was expected. Check the list of parameters to verify that youhave used a valid parameter type. Example: DISP:TEXT ON230