26.11.2012 Views

ECOPROBE 5 - rs dynamics

ECOPROBE 5 - rs dynamics

ECOPROBE 5 - rs dynamics

SHOW MORE
SHOW LESS

You also want an ePaper? Increase the reach of your titles

YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.

• Although there is a water valve fitted to the probe, it is best to avoid inserting the probe into water that<br />

may accumulate in a hole. If a probe is inadvertently inserted into a ‘wet’ hole, the reading will be close to<br />

zero or much lower than expected, because the water valve does not allow vapor to be drawn in. The<br />

position of the sampling probe has to be vertical. Do not tilt the probe since the water valve may cut off the<br />

air as it is not designed to operate in an off vertical position. When surveying a potentially waterlogged<br />

terrain, use a stick to test each hole for seepage water.<br />

• If a delicate measurement is required, the probe may be eliminated. Use only plastic tubes with a dust filter,<br />

in which case always ensure that there is no possibility of liquid being drawn into the instrument.<br />

Heavily contaminated soil vapor may leave a hydrocarbon residue in the entire sampling system, and it is<br />

therefore advisable that the probe and its connecting tube receive special care when being cleaned.<br />

Cleaning the probe<br />

The sampling probe must be cleaned from residual gas and condensation if contamination was found at a<br />

station. Cleaning the probe should be carried out using fresh, non-polluted air. The higher the concentration<br />

and the heavier the hydrocarbons at the measured station, the higher will be the volume of air needed for<br />

cleaning. The probe cleaning can be minimized in the case of soil gas containing contaminants originally<br />

present in a gas phase such as methane. On the contrary, soil gas containing abundant vapor of heavy oils<br />

demands greater volume of cleaning air. Clean air can be blown through by means of a manual pumping<br />

device such as that used for inflating rubber mattresses. It is important not to blow into the sampling probe<br />

with your breath as this may cause moisture to enter the probe and may give incorrect measurements.<br />

Attention needs to be paid to the sampling apertures at the bottom of the sampling probe, which has to be<br />

cleaned of residual soil after removal from the hole.<br />

At some contaminated localities there may be a floating contaminant layer at the ground water table. If the<br />

sampling probe is inserted directly into the pollutant, the probe will be contaminated. In order to avoid false<br />

readings afterwards, immediately clean out the entire probe and pumping inlet with a household dishwashing<br />

detergent and then use clean water for rinsing. Never use organic based detergents since their vapor can<br />

affect the measured values and cause incorrect results.<br />

Connecting tubes<br />

MEASURING – STANDARD GRID<br />

Special care is needed with the plastic connecting tubes between the probe and apparatus. There is an effect<br />

of diffusion of hydrocarbons into plastic. The quality of the plastic materials varies. For example, if a plastic<br />

tube of average quality is used to measure a contaminant concentration higher than 10 000 ppm, the residual<br />

contamination could contribute some tens or hundreds ppm towards the next measurement. The wo<strong>rs</strong>t type,<br />

however, is the soft white silicon tube, while the best is transparent colorless or blue or slightly green plastic<br />

tubes intended for medical purposes. All black rubber tubes are to be avoided. Careful cleaning of the tubes<br />

together with the sampling probe is recommended. When measuring a delicate locality with dramatic shifts<br />

between very high and very low concentrations from one station to another, the best solution is to use a set of<br />

tubes equal to the expected number of stations to be measured within one day. At each station a clean tube is<br />

to be applied. Used tubes usually self-clean if placed on a room heater overnight. The next day they are ready<br />

to be used again.<br />

It is a good practice to set up a locality for “warming up” the instrument and for test measurements.<br />

Before starting to measure, the IR unit needs to be stabilised. Measure three “warming up” stations<br />

in the test locality to avoid any incorrect readings at the beginning of the day’s survey. Use this<br />

locality also for checking if plastic tubing, dust filte<strong>rs</strong> etc. are contaminated.<br />

33

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!