HOW CAN WE UNDERSTAND AND INTERPRET THE RESULTS FROM THE IR METHANE AND IR<br />

TOTAL PETROLEUM CHANNELS?<br />

Both IR Methane and IR Total Petroleum channels are calibrated for methane.The results of the IR methane<br />

channel corresponds exactly with the actual methane concentration.<br />

Results of IR Total petroleum channel:<br />

1. If only methane is present and no other hydrocarbons, the results of the methane channel and the<br />

IR Total Petroleum channel show the actual amount of methane.<br />

2. If a mixture of methane and other hydrocarbons is present, the IR Total Petroleum channel measures<br />

the amount of petroleum hydrocarbons and methane. This value is higher than the result recorded by the<br />

methane channel.<br />

3. If only hydrocarbons are present and no methane – there is no reading on the methane channel, the<br />

IR Total Petroleum channel measures the total amount of petroleum hydrocarbons. It is necessary to<br />

take into account that this channel is calibrated for methane and that the measured value does not<br />

reflect the precise quantity of hydrocarbon present. The resulting value is generally 6 – 10<br />

times higher than the actual concentration of hydrocarbons. You can map the area in a relative way or<br />

divide the results approximately by 10. Compare the IR TP value with the PID value at a station<br />

where methane is not present. Ensure that the TP concentration is not above the photo-ionization<br />

breakdown limit.<br />

Since the IR TP channel is calibrated for methane and the PID analyzer analyzer<br />

on Isobutene, you will never get the<br />

same quantitative data from both analyze<strong>rs</strong>. Reading of IR Total Petroleum channel may be 6 - 10 times<br />

higher than the reading of the PID analyzer, calibrated on Isobutylene (see previous paragraph and<br />

previous page). To obtain comparative readings from IR TP channel and PID channel, both analyze<strong>rs</strong><br />

should be calibrated for the same gas!<br />

Important Note:<br />

IR ANALYZER<br />

For the interpretation of soil vapor data it is essential that all available and relevant information<br />

regarding the surveyed terrain is taken into consideration. The most important information<br />

required, comprises inter alia general geology, the local soil profile, soil characteristics,<br />

hydrogeological conditions, site history, dates of spillages, the type and volume(s) of the<br />

contaminant(s) involved, etc. Without these contributions, an interpretation of <strong>ECOPROBE</strong> 5<br />

data may be flawed.<br />