11. CALIBRATION<br />

CALIBRATION<br />

The <strong>ECOPROBE</strong> 5 system software and <strong>ECOPROBE</strong>_VIEW computer software incorporate three advanced<br />

tools for the precise calibration of all analytical channels, thus ensuring high accuracy of all evaluated data.<br />

1. <strong>ECOPROBE</strong> 5’s internal software enables fast Single Point calibration of the PID analyzer, all<br />

channels of IR analyzer, oxygen and pressure analyze<strong>rs</strong> and the probe thermometer.<br />

2. Another way to calibrate the oxygen and pressure analyze<strong>rs</strong> and the probe thermometer is by using<br />

<strong>ECOPROBE</strong>_VIEW_Light/Plus<br />

communication software.<br />

software. For more details see the User Manual for<br />

3. The <strong>ECOPROBE</strong>_VIEW_Plus software advanced calibration tool enables use<strong>rs</strong> to modify the<br />

original factory calibration curve that might be affected by contamination, by dust or by other field<br />

influences. It can be used for PID and all IR measuring channels. For modifying the entire calibration<br />

curve, the user has to obtain several standards of calibration gas with various concentrations. For more<br />

details see <strong>ECOPROBE</strong>_VIEW_Plus User Manual.<br />

11.1 <strong>ECOPROBE</strong> 5’s PID CALIBRATION<br />

<strong>ECOPROBE</strong> 5’s PID analyzer is calibrated for Isobutene (Isobuthylene). Isobutene represents the most<br />

suitable calibration gas which is also used in laboratory gas chromatographs. At the manufacturer, the PID<br />

analyzer is calibrated using several Isobutene standards with different concentrations using the software for<br />

“Advanced calibration” (see the manual to communication software).<br />

As the PID analyzer does not have a reference channel to compensate for dust influence in contrast to IR<br />

analyzer, it has to be calibrated more frequently. For a fast correction of the PID analyzer, the operator can<br />

calibrate it by using the Single Point Calibration procedure with a 100 ppm Isobutene standard that is supplied<br />

as part of <strong>ECOPROBE</strong> 5 calibration kit.<br />

During a field investigating the PID analyzer may become dusty or optical passages may become dirty due to a<br />

deposit of various particles such as aerosol droplets, due to proper operational procedures not being observed<br />

(dust filte<strong>rs</strong> not used), or if the instrument had been operated for a long period of time. A layer of dust and<br />

various other compounds on the surface of optical system causes attenuation of excitation energy (“light<br />

beam”) resulting in a decrease of the measured values. This attenuation is of the multiplicative character and<br />

does not affect the shape of the calibration curve created at the manufacturer or by the customer using<br />

<strong>ECOPROBE</strong>_VIEW_Plus software. It changes only the direction of the existing curve.<br />