ECOPROBE 5 - rs dynamics
ECOPROBE 5 - rs dynamics
ECOPROBE 5 - rs dynamics
You also want an ePaper? Increase the reach of your titles
YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.
OXYGEN ANALYZER CALIBRATION<br />
PRESSURE ANALYZER CALIBRATION<br />
CALIBRATION<br />
� To calibrate the oxygen analyzer you have to place the<br />
instrument outside in fresh air where the oxygen<br />
concentration is 21%, or use an oxygen calibration<br />
standard.<br />
Go to the calibration locality CALIB.lc (Window W.26)<br />
and let the instrument measure two or three points to<br />
allow it to stabilize itself.<br />
� Measure one station in the fresh air without any tubing<br />
connected. After measuring go to the Main Menu,<br />
select CALIBRATION and continue the calibration<br />
procedure as described for the calibration of the PID<br />
analyzer. In window W.31 activate O2 and type in “21”<br />
if the instrument was calibrated with fresh ambient air,<br />
or type in the real value of the O2 calibration standard<br />
used .<br />
� Go to Calibrate and complete the calibration according<br />
to the described procedure.<br />
� No tubing must be connected to the instrument inlet.<br />
Just start measuring – you do not need any standard<br />
since you measure only atmospheric pressure.<br />
� Open the calibration locality CALIB.lc and let the<br />
instrument measure two or three points to allow it to<br />
stabilize itself.<br />
� After measuring, go to the Main Menu (Window W.2),<br />
select CALIBRATION and continue the calibration<br />
procedure as was described for the calibration of the<br />
PID analyzer.<br />
� In the window W.32 activate the pressure channel. To<br />
continue the calibration you have to know the<br />
atmospheric pressure from the nearest meteorological<br />
station. Meteorological services usually provide the<br />
latest values of atmospheric pressure on their web<br />
sites. If you need to calibrate pressure precisely, you<br />
have to recalculate the pressure value for your actual<br />
altitude.<br />
� Type in the obtained value from the meteorological<br />
station as the actual value of atmospheric pressure.<br />
� Go to Calibrate! and complete the calibration according<br />
to the described procedure.<br />
56<br />
== Select channel ==<br />
PID-S: No ppm<br />
Measured = ---.-- ppm<br />
IR –MET: No ppm<br />
Measured = ---.-- ppm<br />
IR – TP: No ppm<br />
Measured = ---.-- ppm<br />
IR – CO2: No ppm<br />
Measured = ---.-- ppm<br />
O2: Yes 21 %<br />
Measured = 20.59 %<br />
Press.: No mb<br />
Measured = 969.56 mb<br />
Temp.: No C<br />
Measured = ---.-- C<br />
Calibrate!<br />
W.31 - Selection of analyzer for oxygen calibration.<br />
Note: For calibrating the oxygen analyzer it is also possible to use a two-point calibration procedure<br />
implemented in <strong>ECOPROBE</strong>_VIEW software. This calibration may also use a standard with zero<br />
concentration of oxygen (nitrogen). In this way a more reliable calibration and compensation for the decay of<br />
the oxygen sensor are obtained.<br />
== Select channel ==<br />
PID-S: No ppm<br />
Measured = ---.-- ppm<br />
IR –MET: No ppm<br />
Measured = ---.-- ppm<br />
IR – TP: No ppm<br />
Measured = ---.-- ppm<br />
IR – CO2: No ppm<br />
Measured = ---.-- ppm<br />
O2: No %<br />
Measured = 20.18 %<br />
Press.: Yes 967.34 mb<br />
Measured = 959.56 mb<br />
Temp.: No 20.5 C<br />
Measured = ---.-- ppm<br />
Calibrate!<br />
W.32 - Selection of analyzer for pressure<br />
calibration.<br />
The calibration constant for temperature, pressure and oxygen may vary in tenths or hundreds of<br />
units.<br />
Another method to calibrate oxygen and pressure analyze<strong>rs</strong> and the probe thermometer is to use<br />
the <strong>ECOPROBE</strong>_VIEW software. The procedure is described in the manual.