Calibration interval depends on the absolute accuracy required and on the dust condition of surveyed terrain.<br />
Even though filte<strong>rs</strong> are used, the PID unit may accumulate dust and the calibration procedure must<br />
be repeated.<br />
Calibration intervals:<br />
CALIBRATION<br />
To extend the “Single Point” calibration and cleaning interval of the PID analytical<br />
unit the operator has to use a dust filter inserted between <strong>ECOPROBE</strong> 5 gas<br />
inlet and the sampling probe. The filter should be changed depending upon the<br />
amount of dust and extent of contamination of the measuring environment. If<br />
heavily contaminated soil gas was drawn into the instrument, the filter should be<br />
changed before starting the next measurement.<br />
To determine the extent to which a filter is contaminated, set up a locality for test<br />
measurements and measure the filter contamination in the routine way. After the<br />
resetting period, connect only the filter by means of a short piece of tubing and<br />
measure. The degree of contamination will indicate if the filter should be replaced.<br />
Filte<strong>rs</strong> are consumable items which can be purchased from service stations e.g.<br />
Shell, or from automotive spares shops.<br />
� From 1 to 6 months if not working in dusty environment and using dust filte<strong>rs</strong>.<br />
� Once a week if working in a dusty environments with dust filte<strong>rs</strong>.<br />
� Every day if precise quantitative values are required.<br />
The <strong>ECOPROBE</strong> 5’s accessories include a calibration kit for PID Single Point calibration consisting of:<br />
� Cylinder with calibration standard -- LINDE 100 ppm Isobutene (Isobutylene and/or<br />
Isobuthylene is the same compound) mixed with synthetic air. Content of the cylinder (2l/150 atm) is<br />
sufficient for about 200 calibrations if the gas is not wasted.<br />
Calibration gas refills may be purchased from any company providing calibration standard gases for gas<br />
chromatography (Air Products, SIAD, LINDE and othe<strong>rs</strong>).<br />
� Calibration valve for releasing the gas from the cylinder.<br />
� LINDE plastic/metal bag and valve for collecting gas from the cylinder.<br />
49