<strong>ECOPROBE</strong> 5 may display for instance the following values:<br />

• 99.4 ppm of Isobutene on PID analyzer, 0 ppm on IR methane channel, 560 ppm on IR TP channel<br />

(Isobutene is recorded on TP channel and not on the methane channel) and 0 ppm CO 2 on IR CO 2<br />

channel. Pressure during the measurement was 959.56 mb, and the thermometer probe was not<br />

connected.<br />

• For calibrating the PID analyzer we will use only the 99.4 ppm PID value.<br />

• Press ESC on the instrument to return to the Main Menu and select the item CALIBRATION to initiate<br />

the calibration procedure. Calibration Menu (Window W.25) will be opened.<br />

• Select Othe<strong>rs</strong> for calibrating the PID/ IR analyze<strong>rs</strong>, O2, pressure or thermometer and use the right<br />

arrow to open the list of measured localities (Window W.5.1) .<br />

• Select the locality with the calibration measurement (CALIB.lc). Window (W.26) for selecting the point<br />

with calibration values will be opened.<br />

CALIBRATION<br />

Select device<br />

Zero CO 2<br />

Othe<strong>rs</strong><br />

W.25 - Calibration Menu<br />

CALIBRATION<br />

� Use up and down arrows to browse the [measured]<br />

points of the locality CALIB.lc. These arrows change the<br />

coordinates and also the corresponding values of the<br />

points. In our case, fi<strong>rs</strong>t two points were used to stabilize<br />

the instrument. The third one (X=0.0, Y=30.0) comprises<br />

the data of the calibration measurement (99.4 ppm on<br />

PID analyzer, 0 ppm on methane channel, 560 ppm on<br />

IR TP channel, 0 ppm of CO 2 on IR CO 2 channel and<br />

pressure 959.56 mb.<br />

Select this point by pressing the right arrow as advised<br />

on the bottom of the instrument screen.<br />

Window W.27.1 will be displayed.<br />

51<br />

SELECT LOCALITY<br />

XXX<br />

AAA<br />

CALIB<br />

W.5.1 - List of Localities<br />

LOCALITY: CALIB.lc<br />

Select point: 00<br />

X: 0.0 Y: 30.0<br />

PID – S: 99.4ppm<br />

methane: 00ppm<br />

TP: 560ppm<br />

CO2: 00ppm<br />

O2: 20.18%<br />

Pressure: 959.56mb<br />

Temperature: -----.--- C<br />

-----------------------------------------------<br />

Up, Down…move<br />

Righ ……select<br />

W.26 - Window for selecting the point with<br />

calibration values (In this case the given point<br />

has coordinates X=0.0, Y=30.0)