<strong>ECOPROBE</strong> 5 may display for instance the following values:<br />
• 99.4 ppm of Isobutene on PID analyzer, 0 ppm on IR methane channel, 560 ppm on IR TP channel<br />
(Isobutene is recorded on TP channel and not on the methane channel) and 0 ppm CO 2 on IR CO 2<br />
channel. Pressure during the measurement was 959.56 mb, and the thermometer probe was not<br />
connected.<br />
• For calibrating the PID analyzer we will use only the 99.4 ppm PID value.<br />
• Press ESC on the instrument to return to the Main Menu and select the item CALIBRATION to initiate<br />
the calibration procedure. Calibration Menu (Window W.25) will be opened.<br />
• Select Othe<strong>rs</strong> for calibrating the PID/ IR analyze<strong>rs</strong>, O2, pressure or thermometer and use the right<br />
arrow to open the list of measured localities (Window W.5.1) .<br />
• Select the locality with the calibration measurement (CALIB.lc). Window (W.26) for selecting the point<br />
with calibration values will be opened.<br />
CALIBRATION<br />
Select device<br />
Zero CO 2<br />
Othe<strong>rs</strong><br />
W.25 - Calibration Menu<br />
CALIBRATION<br />
� Use up and down arrows to browse the [measured]<br />
points of the locality CALIB.lc. These arrows change the<br />
coordinates and also the corresponding values of the<br />
points. In our case, fi<strong>rs</strong>t two points were used to stabilize<br />
the instrument. The third one (X=0.0, Y=30.0) comprises<br />
the data of the calibration measurement (99.4 ppm on<br />
PID analyzer, 0 ppm on methane channel, 560 ppm on<br />
IR TP channel, 0 ppm of CO 2 on IR CO 2 channel and<br />
pressure 959.56 mb.<br />
Select this point by pressing the right arrow as advised<br />
on the bottom of the instrument screen.<br />
Window W.27.1 will be displayed.<br />
51<br />
SELECT LOCALITY<br />
XXX<br />
AAA<br />
CALIB<br />
W.5.1 - List of Localities<br />
LOCALITY: CALIB.lc<br />
Select point: 00<br />
X: 0.0 Y: 30.0<br />
PID – S: 99.4ppm<br />
methane: 00ppm<br />
TP: 560ppm<br />
CO2: 00ppm<br />
O2: 20.18%<br />
Pressure: 959.56mb<br />
Temperature: -----.--- C<br />
-----------------------------------------------<br />
Up, Down…move<br />
Righ ……select<br />
W.26 - Window for selecting the point with<br />
calibration values (In this case the given point<br />
has coordinates X=0.0, Y=30.0)