� Pressing the right arrow will save the<br />

calibration constant and return you to the<br />

Calibration Menu. You can continue calibrating<br />

the other analyze<strong>rs</strong> or return to the Main Menu.<br />

� If you do not want to continue with the<br />

calibration, press ESC or the left arrow to<br />

return to the Main Menu without making any<br />

changes to the previous setting of the<br />

instrument.<br />

CALIBRATION<br />

W.28 - Window with the resulting<br />

calibration constant<br />

53<br />

Calibration Constants<br />

PID – S: k=1.0398<br />

Press > to save<br />

ESC or < to cancel<br />

The value of the calibration constant provides and indication of the condition of the analytical unit. If it<br />

becomes dusty, the constant value increases. For the PID it may vary from 0.1 to 20. If the value exceeds<br />

20, the system will not allow the calibration procedure to be carried out, and the PID analytical unit<br />

will have to be cleaned.<br />

If completion of calibration is not allowed, it may be also due to expired calibration gas. Obtain a refill and<br />

repeat the calibration procedure before having the analytical unit cleaned.<br />

Never repeat the calibration of the same channel or analyzer immediately! It is necessary to measure<br />

at least one point in any locality before doing the calibration procedure again.