� Pressing the right arrow will save the<br />
calibration constant and return you to the<br />
Calibration Menu. You can continue calibrating<br />
the other analyze<strong>rs</strong> or return to the Main Menu.<br />
� If you do not want to continue with the<br />
calibration, press ESC or the left arrow to<br />
return to the Main Menu without making any<br />
changes to the previous setting of the<br />
instrument.<br />
CALIBRATION<br />
W.28 - Window with the resulting<br />
calibration constant<br />
53<br />
Calibration Constants<br />
PID – S: k=1.0398<br />
Press > to save<br />
ESC or < to cancel<br />
The value of the calibration constant provides and indication of the condition of the analytical unit. If it<br />
becomes dusty, the constant value increases. For the PID it may vary from 0.1 to 20. If the value exceeds<br />
20, the system will not allow the calibration procedure to be carried out, and the PID analytical unit<br />
will have to be cleaned.<br />
If completion of calibration is not allowed, it may be also due to expired calibration gas. Obtain a refill and<br />
repeat the calibration procedure before having the analytical unit cleaned.<br />
Never repeat the calibration of the same channel or analyzer immediately! It is necessary to measure<br />
at least one point in any locality before doing the calibration procedure again.