8. MONITORING<br />
<strong>ECOPROBE</strong> 5 can be used to operate automatically as a monitoring station. You can preset paramete<strong>rs</strong> in<br />
Monitoring Station Configuration File (Window W.23) and leave the instrument anywhere to monitor.<br />
8.1 How to enter and measure a monitoring station<br />
� For entering a new monitoring station use LOCALITY MANAGEMENT from the Main Menu (Window W.2)<br />
and then MONITOR to access a list of monitoring stations (Window W.22).<br />
Select . Monitoring Station Configuration File will be opened (Window W.23). Enter the name of the<br />
monitoring station (for instance MNSALEM). After entering the name and making possible changes to the<br />
paramete<strong>rs</strong>, return to the Main Menu by selecting Save Configuration at the end of the Configuration File or<br />
by pressing ◄ and following the instructions on the screen.<br />
The name of the new main monitoring station (MNSALEM) will be displayed after MONITOR in the Main<br />
Menu. All measured data are organized according to date, which improves sorting data from a monitoring<br />
site. Measurements at a particular monitoring station are logged in 24-hour time periods, and the data are<br />
stored according to the day (date) of observation. To accomplish this, each day a new monitoring substation<br />
is created. When you select the item MONITOR MNSALEM in the Main Menu, you will enter a<br />
measuring mode for the sub-station. The name contains only the fi<strong>rs</strong>t two lette<strong>rs</strong> of the main monitoring<br />
station’s name and the current date – for instance MN040804. At the end of the pre-set time interval the<br />
instrument will start measuring and all data for the day will be stored in this monitoring sub-station. The next<br />
day the instrument will automatically create a new sub-station (MN050804). This name will contain again<br />
the fi<strong>rs</strong>t two lette<strong>rs</strong> of the main station and the new date. In this way data are stored for 10 days. Then the<br />
fi<strong>rs</strong>t monitoring sub-station is permanently deleted to create space on the instrument’s disk. To obtain data<br />
from the fi<strong>rs</strong>t day’s monitoring, it is necessary to visit the station and download the data to the computer<br />
within 10 days.<br />
� For entering an existing monitoring station use LOCALITY MANAGEMENT from the Main Menu and then<br />
MONITOR to access a list of monitoring stations (Window W.2). Select the name of the particular<br />
monitoring station. Its Monitoring Station Configuration File (Window W.22) will be opened for possible<br />
changes. If you do not want to make any changes just return to the Main Menu by pressing ◄ and following<br />
the instructions on the screen. The name of the selected monitoring station will be displayed after<br />
MONITOR in the Main Menu. Select this item to start measuring.<br />
SELECT LOCALITY<br />
<br />
MN040804<br />
MP150903<br />
MT171203<br />
W.22 - List of monitoring stations.<br />
<strong>ECOPROBE</strong> 5 OTHER FEATURES<br />
W.23- Monitoring Station Configuration File<br />
43<br />
MONITOR: noname<br />
---------------------<br />
Grid: Floating<br />
Interval: 100 min<br />
---------------------<br />
Group Measurement No<br />
Pump Speed: 1.25 l/mim<br />
Sampl. interva 0.1 s<br />
Reset period: AUT s<br />
Preint. period: 5 s<br />
Integr. period: 20s<br />
Probe Depth: 0.50 m<br />
Save Curr. Val.: No<br />
Temperature: 99 grad<br />
---------------------<br />
PID On/Off: On<br />
Calibration gas: 118 Isobutene<br />
Units: ppm<br />
Evaluation: Maximum<br />
---------------------<br />
InfraRed On/Off: On<br />
Continuous Measurement: On