ECOPROBE 5 - rs dynamics
ECOPROBE 5 - rs dynamics
ECOPROBE 5 - rs dynamics
You also want an ePaper? Increase the reach of your titles
YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.
CALIBRATION<br />
11.3 SINGLE POINT CALIBRATION OF IR ANALYZER<br />
For IR methane, IR total petroleum and IR CO 2 channels the same Single Point calibration procedure as for the<br />
PID unit is valid. The IR analytical unit requires calibration over a much longer time interval as the unit has a<br />
reference channel capable of compensating for dust, moisture, and temperature influences. The standard IR<br />
calibration interval is about 12 months depending on the dust conditions. The calibration gases for IR channels<br />
are not supplied by the manufacturer and it can be purchased directly from companies providing laboratory and<br />
technical gases like Linde, Siad and othe<strong>rs</strong>.<br />
Both IR Methane and IR Total Petroleum channels are calibrated with methane. For their Single Point<br />
calibration use a gas comprising 10 000 ppm methane mixed with nitrogen.<br />
For calibrating the CO 2 channel use a gas comprising 10 000 ppm CO 2 mixed with nitrogen.<br />
Reserve one calibration bag for calibrating the IR Methane and IR Total Petroleum channels and a second bag<br />
for calibrating the CO 2 channel. Do not interchange them as they may contain residues of calibration gases<br />
and thus cause inaccurate results.<br />
For calibrating all IR channels simultaneously, it is possible to use a mixture of 10 000 ppm of methane<br />
and 10 000 ppm of CO2 (mixture with nitrogen). In this case only one calibration bag is needed for this<br />
mixture of calibration gases.<br />
The calibration bags are not cheap and they should be kept in good order. Do not crease them -- they may lose<br />
their gas-holding capacity.<br />
11.4 Single Point Calibration Procedure of IR channels<br />
The procedure is similar to calibrating the PID analyzer. At<br />
the calibration locality measure the calibration standard -- in<br />
our case a mixture of 10 000 ppm methane and 10 000 ppm<br />
CO2). Then go to CALIBRATION in the Main Menu.<br />
Continue the calibration procedure as described above. Find<br />
the point with measured standard, and in the window W.29<br />
activate the IR channels and type in the actual values of<br />
calibration standard.<br />
Go to Calibrate! and complete the calibration according the<br />
described procedure for the PID analyzer.<br />
The calibration constant for methane may vary from 0.3 to 5;<br />
for total petroleum channel from 0.1 to 30, and for CO2<br />
channel from 0.5 to 2.<br />
54<br />
== Select channel ==<br />
PID-S: No ppm<br />
Measured = ---.-- ppm<br />
IR –MET: Yes 10 000 ppm<br />
Measured = 9 783 ppm<br />
IR – TP: Yes 10 000 ppm<br />
Measured = 9 953 ppm<br />
IR – CO2: Yes 10 000 ppm<br />
Measured = 9 876 ppm<br />
O2: No %<br />
Measured = 18.78 %<br />
Press.: No mb<br />
Measured = 959.56 mb<br />
Temp.: No C<br />
Measured = ---.-- C<br />
Calibrate!<br />
W.29 - Selection of IR channels for calibrating the<br />
IR analyzer.<br />
For calibrating CO2 go directly to the CALIBRATION item in the Main Menu without measuring a calibration<br />
locality; select ZERO CO2 in the Calibration Menu and follow the instructions on the screen. This<br />
procedure is an exception for zero CO2.