13.07.2015 Views

Orion AquaMate 7000 Vis Orion AquaMate 8000 UV-Vis - Daigger

Orion AquaMate 7000 Vis Orion AquaMate 8000 UV-Vis - Daigger

Orion AquaMate 7000 Vis Orion AquaMate 8000 UV-Vis - Daigger

SHOW MORE
SHOW LESS

You also want an ePaper? Increase the reach of your titles

YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.

20 Concentration Measurements—Standard Curve ApplicationMeasuring the Standards for a Standard Curve3. Set Entry concentration and press Enter.4. Press Measure Standards.The Standards screen shows the absorbance of each standard, along with the slope,intercept and correlation coefficient of the standard curve.To measure standards manually (using Manual 6 or Single Cell Holder)1. Install the blank and standards.2. When the parameters are correct, press Run Standards.3. Set Entry concentration.4. Press Measure Blank.If a 6-Position Cell Holder is installed, it moves to the B position to measure the blankand returns to its previous position.5. Press Measure Standards.Thermo Scientific <strong>Orion</strong> <strong>AquaMate</strong> <strong>7000</strong> <strong>Vis</strong> and <strong>Orion</strong> <strong>AquaMate</strong> <strong>8000</strong> <strong>UV</strong>-<strong>Vis</strong> 93

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!