10 Water Analysis ApplicationRunning a Water TestRunning a Water TestWhen a method has been loaded, the following screen will appear.Press the Run Test function key. The following screen appears.Open the sample compartment. If the desired sample holder is not in the beam, press thebutton on the carousel control of the keypad corresponding to the correct cell holder shape(circle for vials or test-tubes, square for 10 mm square cuvettes, rectangle for long pathrectangular cuvettes). If necessary, the instrument will rotate the carousel to place the correctcell holder in the measurement position.Insert a cuvette, tube or vial containing the blank or zero solution into the sample holder.Note Ideally, the same cuvette or vial, or one that has been matched to the sample vial,should be used.See Using a Reverse Color Method for use of reagent blanks in reverse color methods.Thermo Scientific <strong>Orion</strong> <strong>AquaMate</strong> <strong>7000</strong> <strong>Vis</strong> and <strong>Orion</strong> <strong>AquaMate</strong> <strong>8000</strong> <strong>UV</strong>-<strong>Vis</strong> 39