5 Cell CorrectionCell Correction7. Press Run Corr. to start Cell Correction.If you selected Discrete nms mode, first specify the wavelengths using the followingprocedures, and then Cell Correction.Cell Correction measures the other cells against the blank and records, stores and datesthe measurements. From these measurements Cell Correction establishes the requiredcorrection factors, which then are automatically applied during all subsequent tests (ifCell Correction is activated).Specifying Wavelengths for Discrete nms ModeTo specify wavelengths for Discrete nms mode1. Highlight Sample Positioner and press Enter to set this parameter to either Auto 3(when using three large cell holders) or Auto 6 (when using six small cell holders).2. Highlight Number of Matched Cuvettes and press Enter.3. Specify the number of cells you are matching and press Enter.4. Press Set nms to select the wavelengths for which Cell Correction will be run.A list of wavelengths appears.Thermo Scientific <strong>Orion</strong> <strong>AquaMate</strong> <strong>7000</strong> <strong>Vis</strong> and <strong>Orion</strong> <strong>AquaMate</strong> <strong>8000</strong> <strong>UV</strong>-<strong>Vis</strong> 23