5Cell CorrectionEvery test setup screen provides access to Cell Correction.Note Cell Correction is not active from the Main (Basic ATC) screen.Note Cell Correction is active only when the 6-Position Cell Holder is set to eitherAuto 6 or Auto 3. The feature is not active when the cell holder is set to 1-Cell or Manual6, nor when the Single Cell Holder is installed.Before running Cell Correction:• Clean the inside and outside of all the cells to be matched.• Fill the cells with distilled water (or other blank solution), and place them in the samplecompartment (see “Selecting and Positioning Cuvettes” on page 5). Place the blankcuvette in Cell “B” of the cell changer.Cell Correction To run Cell Correction1. Load the test type or stored test.2. If Cell Correction is not visible, highlight More parameters and press Enter.Thermo Scientific <strong>Orion</strong> <strong>AquaMate</strong> <strong>7000</strong> <strong>Vis</strong> and <strong>Orion</strong> <strong>AquaMate</strong> <strong>8000</strong> <strong>UV</strong>-<strong>Vis</strong> 21