4 Sample Positioner SettingManual 6Manual 6Measure a blank and up to five samples without changing cuvettes in the cell changer, usingthe cell position buttons to advance the cell changer to the appropriate position for the nextmeasurement. Place the blank in the blank position and your samples in the other cellpositions. Regardless of where the cell holder is positioned, when you press Measure Blankthe cell holder automatically goes to the blank position and measures the blank. However, youcan use the cell position buttons to select a different position for the measurement.Note When you have the Cell Changer installed, the instrument always considers thematerial in the B position as a blank. This means that even after measuring your blank thefirst time, you can place samples only in positions 1 through 5.20 <strong>Orion</strong> <strong>AquaMate</strong> <strong>7000</strong> <strong>Vis</strong> and <strong>Orion</strong> <strong>AquaMate</strong> <strong>8000</strong> <strong>UV</strong>-<strong>Vis</strong> Thermo Scientific