15ScanningThe wavelength scanning application lets you measure the absorption or percent transmissionspectrum of a sample. You can use scans to determine peak wavelengths or to evaluate thequality of a material.Use the Scanning application for:• Recalling a Test• Setting Up Test Parameters• Cell Correction• Collecting a Baseline Scan• Scanning a Sample• Viewing and Manipulating Scan Data• Rescaling Graphical Scan Data• Determining Peak Height Using a 3-point Net Equation• Calculating the Area Under a Curve• Labeling Peaks and ValleysNote The Scanning application lets you measure only one sample at a time. Auto 6,Auto 3 and Manual 6 are not available for scanned measurements.Note To set a baseline expiration time, press Utility and then highlight BaselineExpiration. Press Enter and set the desired time.To get started, press Test, highlight Scanning and press Enter.Recalling a Test To recall a test1. Press Stored Tests.2. Highlight the test you want to recall and press Enter.Thermo Scientific <strong>Orion</strong> <strong>AquaMate</strong> <strong>7000</strong> <strong>Vis</strong> and <strong>Orion</strong> <strong>AquaMate</strong> <strong>8000</strong> <strong>UV</strong>-<strong>Vis</strong> 65