18 Absorbance DifferenceRecalling a TestRecalling a TestTo recall a test1. In the Absorbance Ratio screen (see “Using the Absorbance Ratio Screen” on page 84),press Stored Tests.2. Highlight the test to recall and press Enter.Using the Absorbance Ratio ScreenUse this screen for:• Setting Up Test Parameters• Cell Correction• Saving a Test• Viewing the list of stored tests• Measuring a Blank• Measuring SamplesSetting Up Test ParametersTo set up test parameters1. In the Absorbance Ratio screen, highlight the parameter to set.2. When the parameters are set, press Save Test to save the test or Run Test to measure ablank or samples.Note If Cell Correction is ON, you must run the Setup Correction application beforeyou can access Run Test or Measure Samples.Note If Auto Save Data is ON, you must enter a Data File Name before you canaccess Run Test or Measure Samples.84 <strong>Orion</strong> <strong>AquaMate</strong> <strong>7000</strong> <strong>Vis</strong> and <strong>Orion</strong> <strong>AquaMate</strong> <strong>8000</strong> <strong>UV</strong>-<strong>Vis</strong> Thermo Scientific