13.07.2015 Views

Orion AquaMate 7000 Vis Orion AquaMate 8000 UV-Vis - Daigger

Orion AquaMate 7000 Vis Orion AquaMate 8000 UV-Vis - Daigger

Orion AquaMate 7000 Vis Orion AquaMate 8000 UV-Vis - Daigger

SHOW MORE
SHOW LESS

Create successful ePaper yourself

Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.

10 Water Analysis ApplicationUsing a Reverse Color MethodProgramming a Reverse Color MethodTo program a new reverse color test, go to the main water test menu, scroll down to theReverse Color option and press the Enter key. The value for Reverse Color will change to On.Scroll down to More Parameters… and press the Enter key.Set the factor (or polynomial equation) using the values supplied with the test kit orcalculated by your laboratory during development of the test.48 <strong>Orion</strong> <strong>AquaMate</strong> <strong>7000</strong> <strong>Vis</strong> and <strong>Orion</strong> <strong>AquaMate</strong> <strong>8000</strong> <strong>UV</strong>-<strong>Vis</strong> Thermo Scientific

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!