22 Performance VerificationNoise2. Press Delete Std.Running the TestIn the Photometric Accuracy screen, make sure the wavelengths, absorbance (or %T) valuesand tolerances are set correctly.To run the Photometric Accuracy testPress Start Test.The results indicate pass or fail for each wavelength.If the test fails, follow these guidelines:• Repeat the test twice to verify that the test is failing consistently.• Follow the guidelines provided with the standard reference materials.• For more troubleshooting advice contact our sales or service representative in your area oruse the information at the beginning of this document to contact us.NoiseThis test measures the amount of noise. The <strong>Vis</strong> instrument measures only at 340 nm.All test parameters are determined by instrument specifications and cannot be changed by theuser. When running the noise test, remember:• Perform the 0A measurement with the cell holder empty. Optionally you can perform the2A measurement with a 2A filter.Thermo Scientific <strong>Orion</strong> <strong>AquaMate</strong> <strong>7000</strong> <strong>Vis</strong> and <strong>Orion</strong> <strong>AquaMate</strong> <strong>8000</strong> <strong>UV</strong>-<strong>Vis</strong> 115