15 ScanningViewing and Manipulating Scan Data2. Press Peaks & Valleys to display the Label Peaks and Valleys window.3. Select the type of labels to display and press Enter.The instrument labels the selected items on your scan data plot.Note Up to nine peaks or valleys can be calculated and displayed.Smoothing DataIf your scan shows sampling noise, you can smooth the data with the smoothing function.To smooth dataWith your scan data displayed on the edit graph screen, press Smoothing [On].Determining Peak Height Using a 3-point Net EquationTo determine 3-point net measurements1. With your scan data displayed on the edit graph screen, press Math.2. Press 3-Pt Net.The 3-point net measurement screen shows the cursor options and three cursor lines(designated for the left, center and right wavelengths).Thermo Scientific <strong>Orion</strong> <strong>AquaMate</strong> <strong>7000</strong> <strong>Vis</strong> and <strong>Orion</strong> <strong>AquaMate</strong> <strong>8000</strong> <strong>UV</strong>-<strong>Vis</strong> 71