Orion AquaMate 7000 Vis Orion AquaMate 8000 UV-Vis - Daigger
Orion AquaMate 7000 Vis Orion AquaMate 8000 UV-Vis - Daigger
Orion AquaMate 7000 Vis Orion AquaMate 8000 UV-Vis - Daigger
Create successful ePaper yourself
Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.
10 Water Analysis ApplicationUsing a Reverse Color MethodReverse color methods require that the instrument measure both a clear blank and a reagentblank to properly calibrate the instrument before measuring test samples.Using a Preprogrammed Reverse Color MethodLoad the method from the library. An example of a reverse color method screen is shownbelow.To run the method, follow the sequence of steps below:1. Press the Run Test function key. The following screen appears:2. Place a cuvette, tube or vial containing a clear, colorless blank solution (such as distilledwater) in the measurement position and press the Measure Blank function key. TheMeasure Rgnt Blank option will appear.46 <strong>Orion</strong> <strong>AquaMate</strong> <strong>7000</strong> <strong>Vis</strong> and <strong>Orion</strong> <strong>AquaMate</strong> <strong>8000</strong> <strong>UV</strong>-<strong>Vis</strong> Thermo Scientific