23 MaintenanceReplacing the Tungsten Lamp11. Position the new assembly as shown above, making sure to align the pins correctly.12. Tighten the thumbscrew.13. Close the lamp door.14. Turn on the instrument and press the UTILITY key.15. Press the arrow keys to highlight Lamp Hours and press ENTER.16. Press ENTER to reset the lamp hours to zero.Thermo Scientific <strong>Orion</strong> <strong>AquaMate</strong> <strong>7000</strong> <strong>Vis</strong> and <strong>Orion</strong> <strong>AquaMate</strong> <strong>8000</strong> <strong>UV</strong>-<strong>Vis</strong> 125