22 Performance VerificationResolution (<strong>UV</strong>-<strong>Vis</strong> only)If the test fails, follow these guidelines:• Repeat the test twice to verify that the test is failing consistently.• Contact technical support for more troubleshooting advice.Resolution (<strong>UV</strong>-<strong>Vis</strong> only)This test measures the ability of the spectrophotometer to resolve adjacent features in aspectrum. The test is performed using a 0.02% (v/v) solution of toluene in hexane andrequires a hexane blank.When running the internal standard test, remember that the wavelengths and tolerance valuesare preset and cannot be changed.To run the Resolution test1. Highlight Resolution and press Enter.Ensure that hexane is in the blank position and toluene in hexane is in cell 1.112 <strong>Orion</strong> <strong>AquaMate</strong> <strong>7000</strong> <strong>Vis</strong> and <strong>Orion</strong> <strong>AquaMate</strong> <strong>8000</strong> <strong>UV</strong>-<strong>Vis</strong> Thermo Scientific