21 KineticsMeasuring SamplesParameterDelay TimeInterval TimeMeasure Blank(as function key)DescriptionEnters the time from Test Initiation to first measurement; allows forsample equilibrationAdv.Enters the time between repeated readingsSelects the frequency of zeroing the instrumentOnce or Every Reading2. When the parameters are set, press Save Test to save the test or Run Test to measure ablank or sample.Note If Cell Correction is ON, you must run the Setup Correction application beforeyou can access Run Test or Measure Samples.Note If Auto Save Data is ON, you must enter a Data File Name before you canaccess Run Test or Measure Samples.Measuring SamplesNote If a 6-Position Cell Changer is installed, place the blank in position B and thesample in cell position #1. The instrument always uses cell position #1 to scan the sample.To measure samples1. In the Kinetics screen, press Run Test.2. If a 6-Position Cell Holder is installed, place the blank in position B and the sample inposition #1.Thermo Scientific <strong>Orion</strong> <strong>AquaMate</strong> <strong>7000</strong> <strong>Vis</strong> and <strong>Orion</strong> <strong>AquaMate</strong> <strong>8000</strong> <strong>UV</strong>-<strong>Vis</strong> 101