22Performance VerificationPerformance Verification lets you check the performance of your instrument with these tests:• Wavelength Accuracy - Internal• Wavelength Accuracy - Standard• Wavelength Repeatability (<strong>UV</strong>-<strong>Vis</strong> only)• Resolution (<strong>UV</strong>-<strong>Vis</strong> only)• Photometric Accuracy• Noise• Stray LightRun the appropriate tests regularly and maintain a log of results to help document thereliability of the instrument and indicate potential performance issues.Note If a printer is installed and turned on, the instrument automatically prints testresults. You can also press Print to print another copy of the results.Accessing the Performance Verification Tests To access the Performance Verification tests1. Press Tests.2. Highlight Performance Verification and press Enter.Thermo Scientific <strong>Orion</strong> <strong>AquaMate</strong> <strong>7000</strong> <strong>Vis</strong> and <strong>Orion</strong> <strong>AquaMate</strong> <strong>8000</strong> <strong>UV</strong>-<strong>Vis</strong> 107