Note:<br />
When the device is in storage it should be recharged monthly to maximise battery life.<br />
If the battery is stored for more than one month at a temperature greater than 21 °C (70 °F), or for more<br />
than one or two weeks at a temperature greater than 45 °C (113 °F), the reserve capacity of the battery<br />
may be affected. It will then be necessary to recharge the battery before using it again.<br />
If the ventilator has been in storage for longer than 30 days connect it to an AC power source,<br />
turn on the unit by the I/O switch at the rear of the ventilator, and let it charge for 15 minutes<br />
prior to starting ventilation.<br />
Note:<br />
Fully charge the internal battery prior to disconnecting from AC Power source (“mains”).<br />
The battery should not be stored for more than two years, whatever the conditions.<br />
6-6 User’s <strong>Manual</strong>