To dismiss the Alarm Logs screen manually:<br />
Press the ENTER key when the cursor is on the “Back” line.<br />
The Alarm Logs screen is dismissed automatically:<br />
• After 15 seconds if no keyboard action is detected<br />
• When a High Priority alarm is triggered<br />
3.4 Silencing the Audible Portion of Alarms<br />
Alarms and Troubleshooting<br />
Note:<br />
Only qualified service personnel may access all alarms and events recorded by the ventilator. Qualified<br />
personnel should refer to the <strong>Puritan</strong> <strong>Bennett</strong> <strong>560</strong> Service <strong>Manual</strong> for further information.<br />
You may silence the audible portion of alarms for 60 seconds at a time. This is referred to as the Audio Paused<br />
function.<br />
To silence the audible portion of activated alarms:<br />
Press the ALARM CONTROL key.<br />
• The audible portion of all activated alarms is paused.<br />
• The visual portions (light indicator and message) of activated alarms remain visible.<br />
• The Audio Paused symbol is displayed at the top right of the screen while the audio<br />
pause function is active.<br />
Figure 3-4. Silencing the Audible Portion of Alarms<br />
If several alarms are activated at the same time, pressing the ALARM CONTROL key affects all current<br />
alarms.<br />
The audible portion of activated alarms is automatically reactivated:<br />
• After 60 seconds, if the cause(s) of the alarm(s) persist(s)<br />
• Whenever a new alarm is activated<br />
Note:<br />
If a key is stuck or held down for 45 seconds a keypad alarm will occur.<br />
3.5 Pausing/Resetting Alarms<br />
User’s <strong>Manual</strong> 3-5