Alarms and Troubleshooting<br />

Table 3-1. Overview of Alarms (Continued)<br />

Alarm Message Cause/<strong>Ventilator</strong> Response Priority<br />

HIGH RATE<br />

HIGH VTE<br />

HIGH VTI<br />

INSP FLOW<br />

RESTART/SRVC<br />

INTENTIONAL VENT<br />

STOP<br />

KEYPAD FAULT<br />

RESTART/SRVC*<br />

*IF PERSISTS<br />

RESTART/SRVC<br />

LOW BATTERY<br />

LOW FIO 2<br />

LOW VTE<br />

LOW VTI<br />

NO<br />

PROXIMAL LINE2*<br />

*IF PERSISTS<br />

RESTART/SRVC<br />

Rate measured greater than<br />

Max Rtot set during three<br />

consecutive breaths.<br />

Alarm activation occurs:<br />

• After three consecutive breaths.<br />

Expired tidal volume greater than<br />

Max VTE set during three<br />

consecutive breaths (in double limb<br />

setup).<br />

Alarm activation occurs:<br />

• After three consecutive breaths.<br />

Inspired tidal volume greater than<br />

Max VTI set during three consecutive<br />

breaths in PSV, CPAP, P A/C, P SIMV,<br />

and V SIMV modes.<br />

Alarm activation occurs:<br />

• After three consecutive breaths.<br />

Inspiratory flow is constant<br />

(± 1 lpm) with normal turbine<br />

temperature and speed conditions.<br />

Contact your service representative<br />

for assistance.<br />

Ventilation has been stopped<br />

voluntarily by the caregiver or<br />

patient.<br />

Keyboard key held down for more<br />

than 45 seconds.<br />

Internal battery capacity < 30 min.<br />

or 8%.<br />

The level of oxygen delivered by the<br />

ventilator is below the Min FIO2 level<br />

set.<br />

Expired tidal volume less than Min<br />

VTE set during three consecutive<br />

breaths (in double-limb setup).<br />

Alarm activation occurs:<br />

• After three consecutive breaths.<br />

Inspired tidal volume less than Min<br />

VTI set during three consecutive<br />

breaths in PSV, CPAP, P A/C, P SIMV<br />

and V SIMV modes.<br />

Alarm activation occurs:<br />

• After three consecutive breaths.<br />

Proximal pressure < 0.6 mbar for<br />

100 ms during inspiration phase of<br />

3 rd breath cycle.<br />

<strong>Ventilator</strong> response: Switch to<br />

internal pressure sensor for pressure<br />

measurement.<br />

Audio<br />

Paused<br />

Avail.<br />

Alarm<br />

Paused<br />

Avail.<br />

MP Yes No<br />

MP Yes No<br />

HP Yes No<br />

HP Yes No<br />

HP Yes Yes<br />

HP No No<br />

HP Yes No<br />

MP Yes No<br />

MP Yes No<br />

MP Yes No<br />

MP Yes No<br />

3-12 User’s <strong>Manual</strong>