Table A-10. <strong>Ventilator</strong> Range, Resolution, and Accuracy (Continued)<br />

<strong>Ventilator</strong> Settings Range, Resolution, and Accuracy<br />

Specifications<br />

Pressure support<br />

Range: OFF or 5 mbar to 55 mbar in valve configuration<br />

(P Support)<br />

Range: 6 mbar to 30 mbar in leak in valve configuration<br />

Resolution: 1 mbar<br />

Accuracy: ± (1 mbar + 10%) of P Support + PEEP setting<br />

Default value: 15 mbar<br />

Depends on PEEP when Relative Pressure is set to YES<br />

I:E Ratio (I:E) Range: from 1:1 to 1:4<br />

Resolution: 1/0.1 s<br />

Accuracy: ± 50 ms or 10%, whichever is greater<br />

Default value: 1/2<br />

I/T Ratio (I/T) Range: 20% to 50%<br />

Resolution: 1%<br />

Accuracy: ± 50 ms or 10%, whichever is greater<br />

Default value: 33%<br />

Inspiratory time (Insp Time) Range: 0.3 s to 2.4 s<br />

Resolution: 0.1 s<br />

Accuracy: ± 50 ms or 10%, whichever is greater<br />

Default value: 1.5 s<br />

Depends on: R-Rate, Vt in V SIMV mode<br />

Depends on: R-Rate in P SIMV mode<br />

Respiratory rate (R-Rate) Range: 5 bpm to 60 bpm in V A/C and P A/C modes<br />

1 bpm to 40 bpm in P SIMV and V SIMV modes<br />

Resolution: 1 bpm<br />

Accuracy: ± 1 bpm<br />

Default value: 13<br />

Depends on: Insp Time and Vt in V SIMV mode<br />

Depends on: Insp Time in P SIMV modes<br />

Depends on: Vt in V A/C mode<br />

Inspiratory sensitivity (I Sens) Range: 1P-5<br />

Resolution: 1<br />

Accuracy: NA<br />

Default value: 2<br />

in CPAP, I Sens is set to 2 and is not adjustable<br />

Exhalation sensitivity (E Sens) Range: 5% to 95% of peak flow<br />

Resolution: 5%<br />

Accuracy: ± (4 lpm +10%) of target exhalation flow based<br />

on E Sens within 50ms<br />

Default value: 25%<br />

In CPAP, E Sens is fixed at 25% and is not adjustable.<br />

Ramp (Flow Pattern) Range: Square (SQ), descending ramp (D), sinusoidal (S)<br />

Resolution: N/A<br />

Default value: Descending ramp (D)<br />

In V SIMV, flow pattern is set to square and is not adjustable<br />

User’s <strong>Manual</strong> A-5