28.02.2013 Views

Puritan Bennett 560 Ventilator User's Manual - Covidien

Puritan Bennett 560 Ventilator User's Manual - Covidien

Puritan Bennett 560 Ventilator User's Manual - Covidien

SHOW MORE
SHOW LESS

You also want an ePaper? Increase the reach of your titles

YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.

Installation and Assembly<br />

After assembling, cleaning, or reassembling the patient circuit, and on a daily basis, inspect the<br />

hoses and other components to ensure that there are no cracks or leaks and that all connections<br />

are secure.<br />

To ensure proper performance of the ventilator, use a patient circuit recommended by <strong>Covidien</strong><br />

in this manual; refer to chapter 4, “Installation and Assembly” and Appendix F, “Parts and<br />

Accessories”. The total specified length of the patient circuit tubing as measured from the<br />

ventilator outlet to the ventilator inlet is 1.1 metres (3.6 feet) to 2.0 metres (6.6 feet). The tubing<br />

must conform to all applicable standards and must be fitted with Ø 22 mm terminals that also<br />

conform to all applicable standards. Ensure that both the length and the internal volume of the<br />

patient circuit are appropriate for the tidal volume: a corrugated tube of Ø 22 mm for adult<br />

patients, and a corrugated tube of Ø 15 mm for pediatric patients with a tidal volume lower<br />

than 200 ml.<br />

Adding accessories to the ventilator breathing circuit, such as a humidifier and water trap(s),<br />

may result in a decrease in tidal volume delivered to the patient due to the added compressible<br />

volume of the accessory. Always assure that the patient is receiving the appropriate inspired<br />

volume when altering the breathing circuit configuration.<br />

Users must always possess an additional breathing circuit and exhalation valve while using the<br />

<strong>Puritan</strong> <strong>Bennett</strong> <strong>560</strong> <strong>Ventilator</strong>.<br />

4.4.1 Choosing the Patient Circuit Type<br />

Single limb circuits are used with breathing modes where spirometry measurements are not required, and double<br />

limb circuits are used with breathing modes where spirometry is required. Be sure to choose the appropriate<br />

circuit in the menu preferences; in particular, ensure that Pediatric Circuit Yes/No is set to YES when using<br />

a pediatric circuit (refer to chapter F, “Parts and Accessories”).<br />

For information regarding validated circuits, visit the SolvITSM Center Knowledge Base by clicking the link at<br />

http://www.puritanbennett.com or contact you customer representative.<br />

4.4.2 Installing the Patient Circuit<br />

The patient circuit is mounted depending on the setup of the circuit used and the accessories used.<br />

The following procedures describe the installation of the patient circuit with a humidifier. To add other accessories,<br />

refer to the installation instructions for the specific accessories used.<br />

To connect a single limb circuit with an exhalation valve: (refer to Figure 4-7)<br />

1. Inspect the components of the patient circuit for any signs of damage, such as cracks (which<br />

might cause leakage). Do not use damaged components to assemble the patient circuit.<br />

2. Install the bacteria filter (item 1) on the TO PATIENT outlet port, as shown.<br />

3. Attach one end of the short circuit tubing (item 2) to the bacteria filter (item 1).<br />

4. Attach the other end of the circuit tubing (item 2) to the inlet port of the humidifier (item 3).<br />

5. Place a water trap (item 4) between the outlet port of the humidifier and the inlet of the<br />

exhalation valve (item 5).<br />

6. Ensure the exhalation valve (item 5) is placed as close as possible to the patient.<br />

7. Connect one end of the proximal pressure tubing (item 7) to the proximal pressure port on the<br />

exhalation valve (item 5) and the other end onto the ventilator patient pressure port (item 8).<br />

8. Connect one end of the exhalation valve tubing (item 6) to the exhalation valve port on the<br />

exhalation valve (item 5) and the other end onto the ventilator exhalation valve port (item 9).<br />

9. To protect the exhalation port (as it will not be used in this configuration), place the cap (if<br />

provided with the breathing circuit) over the exhalation port opening (item 10).<br />

User’s <strong>Manual</strong> 4-7

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!