Specifications<br />
Table A-9. Monitored Parameter Tolerances (Continued)<br />
<strong>Ventilator</strong> Parameters Tolerances<br />
Exhalation Tidal Volume (VTE) ± (10 ml + 10%VTE)*VTE<br />
Total Breath Rate (Rtot) ± 1 bpm<br />
I:E Ratio (I:E)<br />
± 50 ms or 10%, whichever is greater<br />
I/T Ratio (I/T)<br />
± 50 ms or 10%, whichever is greater<br />
Inspiratory Time (I Time) ± 100 ms<br />
Exhalation Time (E Time) ± 100 ms<br />
Inspiratory Minute Volume (M VoI) ± (10ml + 10%)<br />
Vt Sigh ± (20ml + 20%)<br />
FIO2 ± (2.5% + 2.5% FIO2 )<br />
Leak ± (3 lpm + 20%)<br />
Apnea Index (AI) ± 1 ev/h<br />
Apnea Time ± 1 s<br />
% Spontaneous (Spont) ± 1 %<br />
a. The <strong>Puritan</strong> <strong>Bennett</strong> <strong>560</strong> <strong>Ventilator</strong> does not have the capability to reduce pressure below the PEEP pressure<br />
during the exhalation phase.<br />
A.6 Range, Resolution, and Accuracy<br />
Table A-10 lists the ranges, resolutions, and accuracies for ventilator settings, alarm settings, and patient data.<br />
Table A-10. <strong>Ventilator</strong> Range, Resolution, and Accuracy<br />
<strong>Ventilator</strong> Settings Range, Resolution, and Accuracy<br />
Mode Range: V A/C, P A/C, V SIMV, P SIMV, PSV, CPAP<br />
Resolution: N/A<br />
Accuracy: N/A<br />
Default value: P A/C<br />
Tidal volume (Vt) Range: 50 mL to 2000 mL<br />
Resolution: 10 mL<br />
Accuracy: ± (10 ml + 10%) of setting<br />
Default value: 500 mL<br />
Depends on: Insp time, R-Rate in V SIMV and P SIMV<br />
Depends on: Rate and I:E (I/T) in V A/C<br />
Peak Inspiratory Pressure<br />
Range: 5 mbar to 55 mbar in valve configuration<br />
(PIP)<br />
Range: 6 mbar to 30 mbar in leak configuration<br />
Resolution: 1 mbar<br />
Accuracy: ± (1 mbar + 10%) of P Control + PEEP setting<br />
Default value: 15 mbar<br />
Depends on PEEP when Relative Pressure is set to YES<br />
Pressure control<br />
Range: 5 mbar to 55 mbar in valve configuration<br />
(P Control)<br />
Range: 6 mbar to 30 mbar in leak configuration<br />
Resolution: 1 mbar<br />
Accuracy: ± (1 mbar + 10%) of P Control + PEEP setting<br />
Default value: 15 mbar<br />
Depends on PEEP when Relative Pressure is set to YES<br />
A-4 User’s <strong>Manual</strong>