Specifications<br />

Table A-9. Monitored Parameter Tolerances (Continued)<br />

<strong>Ventilator</strong> Parameters Tolerances<br />

Exhalation Tidal Volume (VTE) ± (10 ml + 10%VTE)*VTE<br />

Total Breath Rate (Rtot) ± 1 bpm<br />

I:E Ratio (I:E)<br />

± 50 ms or 10%, whichever is greater<br />

I/T Ratio (I/T)<br />

± 50 ms or 10%, whichever is greater<br />

Inspiratory Time (I Time) ± 100 ms<br />

Exhalation Time (E Time) ± 100 ms<br />

Inspiratory Minute Volume (M VoI) ± (10ml + 10%)<br />

Vt Sigh ± (20ml + 20%)<br />

FIO2 ± (2.5% + 2.5% FIO2 )<br />

Leak ± (3 lpm + 20%)<br />

Apnea Index (AI) ± 1 ev/h<br />

Apnea Time ± 1 s<br />

% Spontaneous (Spont) ± 1 %<br />

a. The <strong>Puritan</strong> <strong>Bennett</strong> <strong>560</strong> <strong>Ventilator</strong> does not have the capability to reduce pressure below the PEEP pressure<br />

during the exhalation phase.<br />

A.6 Range, Resolution, and Accuracy<br />

Table A-10 lists the ranges, resolutions, and accuracies for ventilator settings, alarm settings, and patient data.<br />

Table A-10. <strong>Ventilator</strong> Range, Resolution, and Accuracy<br />

<strong>Ventilator</strong> Settings Range, Resolution, and Accuracy<br />

Mode Range: V A/C, P A/C, V SIMV, P SIMV, PSV, CPAP<br />

Resolution: N/A<br />

Accuracy: N/A<br />

Default value: P A/C<br />

Tidal volume (Vt) Range: 50 mL to 2000 mL<br />

Resolution: 10 mL<br />

Accuracy: ± (10 ml + 10%) of setting<br />

Default value: 500 mL<br />

Depends on: Insp time, R-Rate in V SIMV and P SIMV<br />

Depends on: Rate and I:E (I/T) in V A/C<br />

Peak Inspiratory Pressure<br />

Range: 5 mbar to 55 mbar in valve configuration<br />

(PIP)<br />

Range: 6 mbar to 30 mbar in leak configuration<br />

Resolution: 1 mbar<br />

Accuracy: ± (1 mbar + 10%) of P Control + PEEP setting<br />

Default value: 15 mbar<br />

Depends on PEEP when Relative Pressure is set to YES<br />

Pressure control<br />

Range: 5 mbar to 55 mbar in valve configuration<br />

(P Control)<br />

Range: 6 mbar to 30 mbar in leak configuration<br />

Resolution: 1 mbar<br />

Accuracy: ± (1 mbar + 10%) of P Control + PEEP setting<br />

Default value: 15 mbar<br />

Depends on PEEP when Relative Pressure is set to YES<br />

A-4 User’s <strong>Manual</strong>