Alarms and Troubleshooting<br />

Table 3-2. Alarms and Corrective Actions (Continued)<br />

Alarm Message or<br />

Symptom Possible Reason(s) For The Alarm Event Corrective Action(s)<br />

OCCLUSION<br />

CHECK CIRCUIT<br />

*IF PERSISTS<br />

RESTART/SRVC<br />

OCCLUSION<br />

CHECK CIRCUIT<br />

POWER FAULT<br />

RESTART/SRVC<br />

POWER SUPPLY LOSS<br />

(without message)<br />

PRES SENS FLT1<br />

RESTART/SRVC<br />

PROX SENS FLT2<br />

RESTART/SRVC<br />

REMOVE VALVE OR<br />

CHANGE PRES<br />

REMOVE VALVE<br />

CPAP MODE<br />

Patient circuit obstructed.<br />

A non-vented configuration is being used or<br />

the built-in leak in the mask or in the circuit<br />

may be obstructed or insufficient for the<br />

settings. Note that a high patient or backup<br />

respiratory rate may not sufficiently flush<br />

out CO2 in some vented pediatric masks.<br />

Internal problem in the electrical power<br />

supply.<br />

Electrical power supply cut off by the main<br />

switch when ventilation is in progress.<br />

The internal battery that supplies the<br />

ventilator is entirely discharged.<br />

Defective internal pressure sensor.<br />

Defective proximal pressure sensor or<br />

internal leak of the machine.<br />

The ventilation settings are not compatible<br />

with the type of patient circuit used.<br />

The ventilation settings are not compatible<br />

with the type of patient circuit used.<br />

Clean, unblock, and/or properly connect the<br />

patient circuit.<br />

Replace the non-vented circuit with a<br />

vented one. Clean, unblock the mask or the<br />

circuit of the vented system or switch to a<br />

vented system with a larger leak<br />

configuration. Try to reduce patient’s<br />

backup rate if possible.<br />

Restart ventilator to see if alarm clears. If<br />

not, replace the ventilator and call your<br />

customer service representative.<br />

Press the I/O switch to restore electrical<br />

power to the ventilator and allow<br />

ventilation to continue.<br />

To stop ventilation, press the VENTILATION<br />

ON/OFF key for three seconds.Press the<br />

VENTILATION ON/OFF key again to confirm<br />

stop. (refer to chapter 5, “Operating<br />

Procedures”).<br />

Immediately connect the ventilator to an<br />

AC power outlet or an external DC power<br />

source; otherwise, use an alternate device<br />

to ventilate the patient.<br />

Restart ventilator to see if alarm clears. If<br />

not, have a qualified technician replace the<br />

defective component(s) and call your<br />

customer service representative.<br />

Restart ventilator to see if alarm clears. If<br />

not, have a qualified technician replace the<br />

defective component(s) and call your<br />

customer service representative.<br />

Remove exhalation valve to start ventilation<br />

with less than 5 mbar of difference<br />

between PEEP and PIP or<br />

Increase the difference between PEEP and<br />

PIP to a minimum of 5 mbar.<br />

Remove exhalation valve to start CPAP<br />

ventilation.<br />

SOFTWARE VERSION ERROR Incorrect software version detected. Call your customer service representative.<br />

TURB OVERHEAT<br />

RESTART/SRVC Turbine overheated because of blockage<br />

during operation.<br />

UNKNOWN BATTERY<br />

VALVE MISSING CONNECT<br />

VALVE<br />

Internal battery not recognised as a<br />

<strong>Puritan</strong> <strong>Bennett</strong> product battery.<br />

The ventilation settings are not compatible<br />

with the type of patient circuit used.<br />

Ensure lateral and front openings are not<br />

obstructed.<br />

Check air inlet filter.<br />

Restart ventilator to see if alarm clears. If<br />

not, replace the ventilator and call your<br />

customer service representative.<br />

Call your customer service representative.<br />

Connect exhalation valve.<br />

3-24 User’s <strong>Manual</strong>