A.10 Manufacturer’s Declaration<br />
Specifications<br />
The following tables, Table A-20 through Table A-23, contain the manufacturer’s declarations for the ventilator’s<br />
electromagnetic emissions, electromagnetic immunity, and recommended separation distances between<br />
the ventilator and portable and mobile RF communications equipment, as well as a list of compliant cables.<br />
Warning<br />
Portable and mobile RF communications equipment can affect the performance of the<br />
<strong>Puritan</strong> <strong>Bennett</strong> <strong>560</strong> <strong>Ventilator</strong>. Install and use this device according to the information<br />
contained in this manual.<br />
The ventilator should not be used adjacent to or stacked with other equipment, except as<br />
specified in this manual. If adjacent or stacked use is necessary, the ventilator should be<br />
observed to verify normal operation in the configurations in which it will be used.<br />
Table A-20. Electromagnetic Emissions<br />
The <strong>Puritan</strong> <strong>Bennett</strong> <strong>560</strong> <strong>Ventilator</strong> is intended for use in the electromagnetic environment specified<br />
below. The customer or the user of the ventilator should assure that it is used in such an environment.<br />
RF emissions<br />
CISPR 11 / EN 55011<br />
RF emissions<br />
CISPR 11 / EN 55011<br />
Harmonic emissions<br />
IEC / EN 61000-3-2<br />
Voltage fluctuations/<br />
flicker emissions<br />
IEC / EN 61000-3-3<br />
Group 1 The ventilator uses RF energy only<br />
for its internal functions. Therefore,<br />
its RF emissions are very low and are<br />
not likely to cause any interference<br />
in nearby electronic equipment.<br />
Class B The ventilator is suitable for use in all<br />
establishments including domestic<br />
establishments and those directly<br />
Class A<br />
connected to the public low-voltage<br />
power supply network that supplies<br />
Complies<br />
buildings used for domestic<br />
purposes.<br />
User’s <strong>Manual</strong> A-9