Specifications<br />
A.8 USB<br />
Table A-13. USB Memory Device Specifications<br />
Characteristics Supported formats<br />
USB compatibility USB flash memory USB 2.0 or USB 1.1<br />
Memory file format USB 32 bit format (sector size: 512 - 2,048 bytes)<br />
Number of files Maximum 999<br />
USB size 128 MB to 4 GB<br />
Table A-14. Data Transfer Characteristics<br />
<strong>Ventilator</strong> data description Capacity<br />
Trends capacity 86 MB<br />
Events capacity 512 KB or 5,500 events<br />
Monitorings capacity 42 MB / 48 hours<br />
A.9 Pneumatic<br />
Table A-15. Airway Resistances<br />
Inspiratory Exhalation<br />
1.0 mbar at 30 lpm flow ± 0.1 mbar 0.5 mbar at 30 lpm ± 0.1 mbar<br />
3.7 mbar at 60 lpm flow ± 0.1 mbar 1.1 mbar at 60 lpm ± 0.1 mbar<br />
Table A-16. Patient Circuit Resistances a<br />
Adult Double Limb Pediatric Double Limb<br />
2 mbar at 60 lpm flow b<br />
a. Includes exhalation valve<br />
b. Values obtained from the manufacturer’s directions for use.<br />
Table A-17. Air Inlet Resistance (Filter)<br />
1.1 cmH 2O (1.079 mbar) at 30 lpm flow<br />
Table A-18. Oxygen Inlet Specifications<br />
Maximum<br />
pressure<br />
Maximum<br />
flow<br />
50 kPa (7 psi) 15 lpm<br />
Table A-19. Performance Specifications<br />
Working<br />
pressure<br />
5 mbar –<br />
55 mbar<br />
Sound pressure<br />
level<br />
30 dBA (per NF<br />
EN ISO 17510-1<br />
test conditions)<br />
2 mbar at 30 lpm flow<br />
Maximum<br />
pressure limit<br />
Internal<br />
compliance<br />
(ventilator)<br />
Inspiratory<br />
triggering<br />
response time<br />
(Ttr)<br />
60 mbar .0001 l/mbar 100 ms<br />
A-8 User’s <strong>Manual</strong>