Safety Information<br />
The oxygen supply must be shut off when ventilation is interrupted. Before disconnecting the<br />
oxygen hose, allow the ventilator to continue for a few cycles without oxygen to flush the<br />
patient circuit of excess oxygen.<br />
Before connecting the oxygen supply, ensure that the stud on the oxygen connector is<br />
protruding outwards.<br />
Inspect the oxygen coupler before use to ensure it has its black O-ring attached and in good<br />
condition. Do not use an oxygen coupler with a missing, damaged, or worn O-ring.<br />
Warnings Regarding Electromagnetic Interference<br />
The <strong>Puritan</strong> <strong>Bennett</strong> <strong>560</strong> requires special precautions for electromagnetic compatibility and<br />
should be installed and started according to the recommendations found in Appendix A,<br />
“Specifications.” In particular, the use of nearby mobile and portable communications equipment<br />
using radio frequencies, such as mobile telephones or other systems exceeding the levels set in<br />
the IEC 60601-1-2 standard, may affect its operation. Refer to section A.10, “Manufacturer’s<br />
Declaration”.<br />
The use of any accessory other than those specified, with the exception of the power supplies or<br />
cables sold by <strong>Covidien</strong>, may lead to an increase in electromagnetic emissions or a decrease in<br />
the equipment protection against electromagnetic emissions. If the ventilator is used adjacent to<br />
such accessories or stacked with such devices, the ventilator’s performance should be monitored<br />
to verify normal operation.<br />