Table 3-1. Overview of Alarms (Continued)<br />

Alarm Message Cause/<strong>Ventilator</strong> Response Priority<br />

BUZZER FAULT4<br />

RESTART/SRVC<br />

CALIBRATE FIO 2<br />

CALIBRATION FAIL<br />

CHECK BATTERY<br />

CHARGE<br />

IF PERSISTS<br />

RESTART/SRVC<br />

CHECK EXH VALVE*<br />

*IF PERSISTS<br />

RESTART/SRVC<br />

CHECK EXH VALVE<br />

PRESSURE<br />

CHECK FIO 2<br />

SENSOR<br />

CHECK PROXIMAL<br />

LINE1*<br />

*IF PERSISTS<br />

RESTART/SRVC<br />

CHECK REMOTE<br />

ALARM<br />

Buzzer Battery Failure. The Battery<br />

Buzzer Voltage is too low.<br />

Internal technical problem that<br />

prevents the battery sounding the<br />

POWER SUPPLY LOSS alarm.<br />

An FIO 2 sensor is detected and has<br />

not been calibrated.<br />

Failure of one calibration point of<br />

the internal exhaled flow sensor.<br />

Consequence: failed calibration<br />

point is replaced by the default<br />

point.<br />

Internal battery charging failure.<br />

Consequence: charging of the<br />

internal battery impossible.<br />

Inspired tidal volume during<br />

exhalation < 20% of Inspired tidal<br />

volume and Inspired tidal volume ><br />

20mL.<br />

Exhalation valve obstructed.<br />

Internal ventilation fault related to<br />

exhalation valve detection sensor.<br />

(pressure sensor)<br />

FIO 2 measurement is less than 18%.<br />

Recalibrate or change FIO 2 sensor.<br />

1. Loss of signal from the proximal<br />

pressure sensor<br />

Consequence: switch to internal<br />

pressure sensor for the pressure<br />

measurement.<br />

Alarm activation occurs:<br />

In the event of signal loss (1):<br />

After one ventilation cycle<br />

or<br />

In the event of signal loss (2) and<br />

after the 17th breath cycle:<br />

After 17 seconds for P A/C and<br />

V A/C modes, or after the maximum<br />

time between 17 seconds and<br />

Apnea Time + 4 seconds for CPAP,<br />

PSV, P SIMV, and V SIMV modes<br />

No activation of Nurse Call or<br />

remote alarm system when an alarm<br />

is in progress.<br />

Alarms and Troubleshooting<br />

Audio<br />

Paused<br />

Avail.<br />

Alarm<br />

Paused<br />

Avail.<br />

MP Yes Yes<br />

MP Yes Yes<br />

MP Yes Yes<br />

MP Yes Yes<br />

HP Yes No<br />

HP Yes Yes<br />

HP Yes No<br />

MP Yes No<br />

MP Yes Yes<br />

User’s <strong>Manual</strong> 3-9