Table 3-1. Overview of Alarms (Continued)<br />

Alarm Message Cause/<strong>Ventilator</strong> Response Priority<br />

OCCLUSION<br />

CHECK CIRCUIT*<br />

*IF PERSISTS<br />

RESTART/SRVC<br />

OCCLUSION<br />

CHECK CIRCUIT<br />

PATIENT<br />

DISCONNECTION*<br />

*IF PERSISTS<br />

RESTART/SRVC<br />

POWER FAULT<br />

RESTART/SRVC<br />

POWER SUPPLY<br />

LOSS<br />

(no message)<br />

Occurs in VALVE configuration when<br />

the tidal volume is measured below<br />

20ml during three consecutive<br />

breaths for PSV, CPAP, PA/C and P<br />

SIMV modes.<br />

Alarm activation occurs after three<br />

consecutive breaths if the tidal<br />

volume is less than 20 mL.<br />

Occurs in LEAK configuration when<br />

the LEAK level is not sufficient to<br />

flush the CO 2 from patient<br />

exhalation. The built-in LEAK in the<br />

mask may be obstructed.<br />

The built-in leak for the mask is not<br />

sufficient for the settings.<br />

Alarm activation occurs if conditions<br />

remain for the maximum time<br />

between:<br />

• disconnection time and 60/R-<br />

Rate in P A/C and V A/C mode<br />

• disconnection time and (Apnea<br />

time +2 sec) in CPAP and PSV<br />

mode<br />

• disconnection time and (60/R-<br />

Rate + Insp time) in P SIMV and<br />

V SIMV mode.<br />

If the flow is greater than 130 lpm<br />

during the inspiratory phase.<br />

In V A/C and V SIMV modes, if<br />

patient pressure is lower than Min<br />

PIP.<br />

In PSV, CPAP, P A/C modes<br />

and P SIMV if patient pressure is<br />

lower than (P Support +<br />

PEEP) - 20% or (P Control + PEEP) -<br />

20%.<br />

Detection of a fault in the electrical<br />

power supply system.<br />

1. Electrical power supply to the<br />

machine is interrupted with the I/O<br />

switch when ventilation is in<br />

progress<br />

or<br />

2. Battery fully discharged when it<br />

was the only source of power to the<br />

ventilator.<br />

Consequence: ventilation stops<br />

immediately. Ventilation restarts<br />

immediately when the switch is<br />

pressed in case 1 (above) or after<br />

restoration of the AC or DC supply<br />

in case 2 (above).<br />

Alarms and Troubleshooting<br />

Audio<br />

Paused<br />

Avail.<br />

Alarm<br />

Paused<br />

Avail.<br />

HP Yes No<br />

HP Yes No<br />

HP Yes No<br />

MP Yes Yes<br />

VHP No Yes<br />

User’s <strong>Manual</strong> 3-13