2.8 Ventilation Menu<br />

Ventilation menu with ventilation on standby.<br />

Ventilation menu during ventilation.<br />

1<br />

General information line:<br />

Displays the current ventilation mode,<br />

along with the following:<br />

• Battery symbol if the device is<br />

powered by the internal battery.<br />

• Audio paused symbol if an<br />

alarm is currently inhibited.<br />

• Alarm paused symbol if an<br />

alarm has been cancelled manually<br />

and the cause of the alarm<br />

remains.<br />

• Apnea Alarm deactivation<br />

• Exhalation valve symbol.<br />

• No exhalation valve symbol.<br />

• Absolute ABS symbol.<br />

• Relative REL symbol.<br />

4 Bargraph:<br />

Displays pressure generation during<br />

ventilation.<br />

1<br />

2<br />

3<br />

4<br />

1<br />

2<br />

3<br />

4<br />

2<br />

5<br />

Ventilation settings:<br />

Displays the specific ventilation<br />

parameter values for the<br />

currently selected ventilation<br />

mode.<br />

Status/monitored data window:<br />

• Ventilation stopped<br />

(Standby): displays the<br />

message, “PRESS TO<br />

START VENTILATION.”<br />

• Ventilation on: parameters<br />

are monitored and displayed.<br />

• The Inspiratory Effort<br />

Detected symbol appears<br />

adjacent to the monitored I:E<br />

ratio when the patient<br />

actively triggers a breath.<br />

Figure 2-4. Ventilation Menu Display<br />

<strong>Ventilator</strong> Overview<br />

User’s <strong>Manual</strong> 2-7<br />

3<br />

6<br />

Preferences menu access<br />

line:<br />

Highlight this line and press<br />

the ENTER key to<br />

display the Preferences<br />

menu.<br />

Alarm conditions window:<br />

• For Active alarms, scrolls<br />

through active alarm<br />

messages in flashing<br />

reverse video.<br />

• For Inactive alarms,<br />

displays the last alarm<br />

along with its trigger date<br />

and end-of-event time.<br />

Refer to chapter 3, “Alarms<br />

and Troubleshooting” for<br />

details.<br />

5<br />

6<br />

5<br />

6