2.8 Ventilation Menu<br />
Ventilation menu with ventilation on standby.<br />
Ventilation menu during ventilation.<br />
1<br />
General information line:<br />
Displays the current ventilation mode,<br />
along with the following:<br />
• Battery symbol if the device is<br />
powered by the internal battery.<br />
• Audio paused symbol if an<br />
alarm is currently inhibited.<br />
• Alarm paused symbol if an<br />
alarm has been cancelled manually<br />
and the cause of the alarm<br />
remains.<br />
• Apnea Alarm deactivation<br />
• Exhalation valve symbol.<br />
• No exhalation valve symbol.<br />
• Absolute ABS symbol.<br />
• Relative REL symbol.<br />
4 Bargraph:<br />
Displays pressure generation during<br />
ventilation.<br />
1<br />
2<br />
3<br />
4<br />
1<br />
2<br />
3<br />
4<br />
2<br />
5<br />
Ventilation settings:<br />
Displays the specific ventilation<br />
parameter values for the<br />
currently selected ventilation<br />
mode.<br />
Status/monitored data window:<br />
• Ventilation stopped<br />
(Standby): displays the<br />
message, “PRESS TO<br />
START VENTILATION.”<br />
• Ventilation on: parameters<br />
are monitored and displayed.<br />
• The Inspiratory Effort<br />
Detected symbol appears<br />
adjacent to the monitored I:E<br />
ratio when the patient<br />
actively triggers a breath.<br />
Figure 2-4. Ventilation Menu Display<br />
<strong>Ventilator</strong> Overview<br />
User’s <strong>Manual</strong> 2-7<br />
3<br />
6<br />
Preferences menu access<br />
line:<br />
Highlight this line and press<br />
the ENTER key to<br />
display the Preferences<br />
menu.<br />
Alarm conditions window:<br />
• For Active alarms, scrolls<br />
through active alarm<br />
messages in flashing<br />
reverse video.<br />
• For Inactive alarms,<br />
displays the last alarm<br />
along with its trigger date<br />
and end-of-event time.<br />
Refer to chapter 3, “Alarms<br />
and Troubleshooting” for<br />
details.<br />
5<br />
6<br />
5<br />
6