28.02.2013 Views

Puritan Bennett 560 Ventilator User's Manual - Covidien

Puritan Bennett 560 Ventilator User's Manual - Covidien

Puritan Bennett 560 Ventilator User's Manual - Covidien

SHOW MORE
SHOW LESS

Create successful ePaper yourself

Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.

E Alarms Tests<br />

Before connecting the ventilator to the patient, perform the following tests to ensure the ventilator‘s alarms<br />

are working properly.<br />

Warning<br />

Do not perform ventilator alarm tests while the patient is connected to the ventilator. Provide<br />

the patient with an alternate means of ventilation before conducting these tests.<br />

If the ventilator fails any alarm test or if you cannot complete these tests, see the<br />

Troubleshooting section (refer to chapter 3, “Alarms and Troubleshooting”) of this manual or<br />

call your equipment supplier or <strong>Covidien</strong> (refer to section 8.3, “Service Assistance”).<br />

The setting of the Min PIP alarm must be adjusted for the patient, but must also be set high<br />

enough to allow the PATIENT DISCONNECTION alarm to trigger properly. Perform the Low<br />

Pressure Test (refer to section E.1, “Low Pressure Test”) to ensure the Min PIP alarm is properly<br />

set.<br />

Note:<br />

Many of the functions of the ventilator are not accessible when the Locking key is enabled. For<br />

additional assistance contact your Clinician or equipment representative.<br />

Most of these tests require that an approved patient circuit be connected to the ventilator. Ensure that<br />

your patient circuit is properly connected prior to performing these tests.<br />

E.1 Low Pressure Test<br />

Warning<br />

The setting of the Min PIP alarm must be adjusted for the patient, but must also be set high<br />

enough to allow the PATIENT DISCONNECTION alarm to trigger properly. Perform the following<br />

test to ensure the Low PIP alarm is properly set.<br />

1. Before proceeding, set the ventilation and alarm parameters specified by the patient’s clinician<br />

and specify a single or dual circuit setup.<br />

2. Press the VENTILATION ON/OFF key to start ventilation.<br />

3. Keep the patient‘s end of the breathing circuit open and allow ventilation to continue.<br />

4. Wait for (Apnea Time + 2 seconds; Apnea time is not always 5 seconds), then ensure that:<br />

• the High priority indicator (red colour) lights up<br />

• the “PATIENT DISCONNECTION” alarm is displayed<br />

• the audible alarm sounds<br />

5. Press the ALARM CONTROL key once to silence the alarm.<br />

6. Press and hold the VENTILATION ON/OFF key for three (3) seconds, then release it.<br />

Press the VENTILATION ON/OFF key again to confirm stop.The ventilator will switch to<br />

Standby mode and cancel the alarms.

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!