Puritan Bennett 560 Ventilator User's Manual - Covidien
Puritan Bennett 560 Ventilator User's Manual - Covidien
Puritan Bennett 560 Ventilator User's Manual - Covidien
Create successful ePaper yourself
Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.
E Alarms Tests<br />
Before connecting the ventilator to the patient, perform the following tests to ensure the ventilator‘s alarms<br />
are working properly.<br />
Warning<br />
Do not perform ventilator alarm tests while the patient is connected to the ventilator. Provide<br />
the patient with an alternate means of ventilation before conducting these tests.<br />
If the ventilator fails any alarm test or if you cannot complete these tests, see the<br />
Troubleshooting section (refer to chapter 3, “Alarms and Troubleshooting”) of this manual or<br />
call your equipment supplier or <strong>Covidien</strong> (refer to section 8.3, “Service Assistance”).<br />
The setting of the Min PIP alarm must be adjusted for the patient, but must also be set high<br />
enough to allow the PATIENT DISCONNECTION alarm to trigger properly. Perform the Low<br />
Pressure Test (refer to section E.1, “Low Pressure Test”) to ensure the Min PIP alarm is properly<br />
set.<br />
Note:<br />
Many of the functions of the ventilator are not accessible when the Locking key is enabled. For<br />
additional assistance contact your Clinician or equipment representative.<br />
Most of these tests require that an approved patient circuit be connected to the ventilator. Ensure that<br />
your patient circuit is properly connected prior to performing these tests.<br />
E.1 Low Pressure Test<br />
Warning<br />
The setting of the Min PIP alarm must be adjusted for the patient, but must also be set high<br />
enough to allow the PATIENT DISCONNECTION alarm to trigger properly. Perform the following<br />
test to ensure the Low PIP alarm is properly set.<br />
1. Before proceeding, set the ventilation and alarm parameters specified by the patient’s clinician<br />
and specify a single or dual circuit setup.<br />
2. Press the VENTILATION ON/OFF key to start ventilation.<br />
3. Keep the patient‘s end of the breathing circuit open and allow ventilation to continue.<br />
4. Wait for (Apnea Time + 2 seconds; Apnea time is not always 5 seconds), then ensure that:<br />
• the High priority indicator (red colour) lights up<br />
• the “PATIENT DISCONNECTION” alarm is displayed<br />
• the audible alarm sounds<br />
5. Press the ALARM CONTROL key once to silence the alarm.<br />
6. Press and hold the VENTILATION ON/OFF key for three (3) seconds, then release it.<br />
Press the VENTILATION ON/OFF key again to confirm stop.The ventilator will switch to<br />
Standby mode and cancel the alarms.