A.3 Indicators and Alarms<br />
Table A-5. Power Indicators<br />
Ventilation ON/OFF AC power DC power Internal Battery<br />
• Blue in standby mode<br />
• Not lit if ventilation is in<br />
progress.<br />
Table A-6. Alarm Indicators<br />
High Priority Medium Priority<br />
Red flashing LED Yellow flashing LED<br />
Table A-7. Audio Alarms<br />
Audio<br />
Paused<br />
Alarm Volume<br />
60 s ± 1 s 65 to 85 dBA ± 10% at 1 meter<br />
A.4 Performance<br />
A.4.1 Specifications<br />
A.5 Monitored Parameters<br />
Green Green • Flashing if the battery<br />
charge is in progress.<br />
• Continuously lit if the<br />
ventilator is powered by<br />
the internal battery.<br />
Table A-8. Performance Parameter Specifications and Tolerances<br />
Settings Range Tolerances<br />
Volume 50 to 2000 ml ± (10 ml + 10%)<br />
Pressure 5 to 55 mbar ± (1 mbar +10%)<br />
Time 0.3 to 2.4 s ± 50 ms or 10%, whichever is<br />
greater<br />
Rate 1 to 60 bpm ± 1 bpm<br />
Inspiratory Sensitivity 1 P to 5 N/A<br />
Exhalation Sensitivity 5 to 95% ± (4 lpm + 10%) of target<br />
exhalation flow based on E Sens<br />
within 50ms<br />
Vt Sigh Vt x1 to Vt x 2 ± (20ml + 20%)<br />
I:E 1:4 to 1:1 ± 50 ms or 10%, whichever is<br />
greater<br />
I/T 20% to 50% ± 50 ms or 10%, whichever is<br />
greater<br />
Table A-9. Monitored Parameter Tolerances<br />
<strong>Ventilator</strong> Parameters Tolerances<br />
Peak Inspiratory Pressure (PIP) ± (2 mbar + 8%)<br />
Positive End Expiratory Pressure (PEEP) a<br />
± (2 mbar + 8%)<br />
Inspiratory Tidal Volume (VTI) ± (10 ml + 10%VTI)*Rate<br />
Specifications<br />
User’s <strong>Manual</strong> A-3