Puritan Bennett 560 Ventilator User's Manual - Covidien
Puritan Bennett 560 Ventilator User's Manual - Covidien
Puritan Bennett 560 Ventilator User's Manual - Covidien
You also want an ePaper? Increase the reach of your titles
YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.
Installation and Assembly<br />
To connect the DC power cable to the ventilator:<br />
1. Line up the red marker dot on the ventilator’s DC power connector with the marker on the DC<br />
power cable (Figure 4-6, item 1).<br />
Figure 4-6. Connecting the DC Power Cable to the <strong>Ventilator</strong><br />
2. Push the DC power cable onto the ventilator’s DC power connector (Figure 4-6, item 2).<br />
• You will hear a locking “click”.<br />
• The DC POWER indicator on the top left corner of the ventilator illuminates (see<br />
Figure 4-4).<br />
To disconnect the DC power cable from the ventilator, slide the locking ring (Figure 4-6, item 3) back and pull<br />
the plug away from the ventilator’s rear panel to disengage it.<br />
An “DC POWER DISCONNECTION“ alarm signals an automatic switch to the internal battery in case the external<br />
DC power source fails or becomes disconnected.<br />
4.4 Patient Circuit<br />
1<br />
Align the red<br />
markers (dots and/or<br />
lines) before<br />
connecting<br />
the cable.<br />
Push in to Connect the<br />
DC Power Cable<br />
Warning<br />
Before opening the packaging for the Patient Circuit, ensure that no damage is evident to the<br />
packaging or its contents. Do not use if evidence of damage exists.<br />
For pediatric use, ensure that the patient circuit type fits, and, in all respects, is suitable for use<br />
with a child. Use a pediatric circuit for patients that weigh under 53 lb. (23 kg). See Table F-2,<br />
List of Circuits, on page F-2 for a list of recommended patient circuits.<br />
If exhaled tidal volume measurements are required to ensure correct patient ventilation a<br />
double limb patient circuit configuration must be used in order to detect leaks. In this case, both<br />
the minimum and maximum VTE alarm parameters must be properly set to warn in the event of<br />
patient disconnection.<br />
The patient circuit should always be positioned to avoid hindering the patient's movements, to<br />
prevent accidental disconnection or leakage, and to minimise the risk of patient strangulation.<br />
Ensure that the ventilator’s immediate surroundings allow for the proper operational<br />
connection of the device without folding, pinching, or damaging any of the required cables or<br />
tubes, and that the connection of the patient circuit to the patient provides for a secure,<br />
comfortable fit.<br />
The patient circuit is intended for single use by a single patient and should be changed<br />
according to the manufacturer’s recommendations and according to the patient circuit lifetime.<br />
Refer to the instructions for use supplied by the manufacturer of the patient circuit (included<br />
with the ventilator) and chapter 4, “Installation and Assembly”.<br />
4-6 User’s <strong>Manual</strong><br />
2<br />
3