28.02.2013 Views

Puritan Bennett 560 Ventilator User's Manual - Covidien

Puritan Bennett 560 Ventilator User's Manual - Covidien

Puritan Bennett 560 Ventilator User's Manual - Covidien

SHOW MORE
SHOW LESS

You also want an ePaper? Increase the reach of your titles

YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.

Installation and Assembly<br />

• Connected to the FROM PATIENT port:<br />

This filter protects the internal exhalation flow sensor from the gases exhaled by the patient.<br />

Refer to Figure 4-8, item 10.<br />

Figure 4-12. Bacteria Filter<br />

Refer to the manufacturer’s instructions for more information about the use and maintenance of the bacteria<br />

filter(s).<br />

4.6 Humidifier<br />

The humidifier (Figure 4-13) adds moisture (water vapour) and warms the gas in the patient circuit. It is<br />

inserted into the patient circuit between the main outlet and the patient (refer to Figure 4-7 and Figure 4-8).<br />

Warning<br />

During invasive ventilation (when an artificial airway bypasses the patient’s upper respiratory<br />

system), the patient’s upper respiratory system cannot humidify the incoming gas. For this<br />

reason, the use of a humidifier, to minimise drying of the patient’s airways and subsequent<br />

irritation and discomfort, must be used.<br />

Always position a humidification device so that it is lower than both the ventilator and the<br />

patient. Use water traps, if necessary, to limit water in the patient circuit and periodically empty<br />

these water traps.<br />

If a heated humidifier is used, you should always monitor the temperature of the gas delivered<br />

to the patient. Gas delivered from the ventilator that becomes too hot may burn the patient's<br />

airway.<br />

Adding accessories to the ventilator breathing circuit, such as a humidifier and water trap(s),<br />

may result in a decrease in tidal volume delivered to the patient due to the added compressible<br />

volume of the accessory. Always assure that the patient is receiving the appropriate inspired<br />

volume when altering the breathing circuit configuration.<br />

Figure 4-13. Humidifier<br />

When a humidification device is used, any condensation that forms in the patient circuit is collected in the<br />

4-12 User’s <strong>Manual</strong>

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!