Puritan Bennett 560 Ventilator User's Manual - Covidien
Puritan Bennett 560 Ventilator User's Manual - Covidien
Puritan Bennett 560 Ventilator User's Manual - Covidien
You also want an ePaper? Increase the reach of your titles
YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.
Installation and Assembly<br />
• Connected to the FROM PATIENT port:<br />
This filter protects the internal exhalation flow sensor from the gases exhaled by the patient.<br />
Refer to Figure 4-8, item 10.<br />
Figure 4-12. Bacteria Filter<br />
Refer to the manufacturer’s instructions for more information about the use and maintenance of the bacteria<br />
filter(s).<br />
4.6 Humidifier<br />
The humidifier (Figure 4-13) adds moisture (water vapour) and warms the gas in the patient circuit. It is<br />
inserted into the patient circuit between the main outlet and the patient (refer to Figure 4-7 and Figure 4-8).<br />
Warning<br />
During invasive ventilation (when an artificial airway bypasses the patient’s upper respiratory<br />
system), the patient’s upper respiratory system cannot humidify the incoming gas. For this<br />
reason, the use of a humidifier, to minimise drying of the patient’s airways and subsequent<br />
irritation and discomfort, must be used.<br />
Always position a humidification device so that it is lower than both the ventilator and the<br />
patient. Use water traps, if necessary, to limit water in the patient circuit and periodically empty<br />
these water traps.<br />
If a heated humidifier is used, you should always monitor the temperature of the gas delivered<br />
to the patient. Gas delivered from the ventilator that becomes too hot may burn the patient's<br />
airway.<br />
Adding accessories to the ventilator breathing circuit, such as a humidifier and water trap(s),<br />
may result in a decrease in tidal volume delivered to the patient due to the added compressible<br />
volume of the accessory. Always assure that the patient is receiving the appropriate inspired<br />
volume when altering the breathing circuit configuration.<br />
Figure 4-13. Humidifier<br />
When a humidification device is used, any condensation that forms in the patient circuit is collected in the<br />
4-12 User’s <strong>Manual</strong>