To connect the oxygen supply system to the ventilator:<br />

Refer to Figure 4-16 as required:<br />

Installation and Assembly<br />

1. Inspect the oxygen supply’s connector (Figure 4-16, item 1) to ensure that connector’s<br />

black O-ring (item 2) is not missing.<br />

2. Push the oxygen supply’s oxygen connector (item 1) into the ventilator’s oxygen connector<br />

(Figure 4-16, item 3).<br />

• the ventilator’s oxygen connector’s locking stud (item 4) retracts.<br />

• the ventilator’s oxygen connector’s locking tab (item 5) is released, ensuring that the oxygen<br />

connection is locked and secured in place.<br />

1<br />

2 3<br />

Figure 4-16. Connecting the Oxygen Supply System<br />

To disconnect the oxygen supply system from the ventilator:<br />

Note:<br />

Ensure the oxygen source is turned off prior to placing the ventilator in Standby or turning off the ventilator.<br />

1. Stop the oxygen flow from the oxygen supply.<br />

2. Press the locking tab of the ventilator’s oxygen connector, as shown in Figure 4-17, to unlock<br />

the oxygen connection.<br />

Figure 4-17. Disconnecting the Oxygen Supply System<br />

3. Disconnect the oxygen supply’s oxygen connector by pulling it towards you.<br />

The ventilator’s oxygen connector’s locking stud (Figure 4-16, item 4) will then extend<br />

outwards,<br />

which is required before the oxygen connector can be reconnected.<br />

User’s <strong>Manual</strong> 4-15<br />