Installation and Assembly<br />

The <strong>Puritan</strong> <strong>Bennett</strong> <strong>560</strong> <strong>Ventilator</strong> can be used with an optional oxygen analyser with<br />

minimum and maximum concentration alarms. Always measure the delivered oxygen with a<br />

calibrated oxygen analyser (FIO 2 kit) that features a minimum and maximum concentration<br />

alarm in order to ensure that the prescribed oxygen concentration is delivered to the patient.<br />

Oxygen administered to the patient is introduced from an external source into the machine through the oxygen<br />

connector at the rear of the ventilator. It is then integrated into the total volume of delivered gas. Remove<br />

the oxygen inlet connector from the back of the ventilator when external oxygen is not in use.<br />

The specific oxygen flow to the patient depends on the physiological characteristics of the patient and the<br />

ventilator settings.<br />

The oxygen flow setting should be adjusted for each patient and established in relation to a calibrated oxygen<br />

monitor measurement. Since the factors that affect administered oxygen flow may change over time, you<br />

must ensure that these settings always correspond to the current oxygen therapy objectives specified by the<br />

physician.<br />

4.8.2 Connecting the Oxygen Supply<br />

Warning<br />

Ensure that the only gas supplied to the ventilator through the dedicated oxygen supply<br />

connector is medical-grade oxygen.<br />

The hose connecting the ventilator to the oxygen source must be designed exclusively for use<br />

with medical-grade oxygen. Under no circumstances should the oxygen hose be modified by<br />

the user. In addition, the hose must be installed without the use of lubricants.<br />

A connector (Figure 4-15, item 1) for an external low pressure oxygen source is available at the rear of the ventilator.<br />

It is essential that you also use the special coupler (item 2) supplied with the ventilator to attach the<br />

external low pressure oxygen source to the connector. The connector is also fitted with a non-return airtight<br />

valve system. The non-return airtight valve system includes a stud (item 3) and a locking tab (item 4).<br />

1<br />

Warning<br />

3<br />

Figure 4-15. Rear Panel Oxygen Connector<br />

Before connecting the oxygen supply, ensure that the stud on the oxygen connector<br />

(Figure 4-15, item 3) is protruding outwards.<br />

Inspect the oxygen coupler (Figure 4-16, item 2) before use to ensure it has its black O-ring<br />

attached and in good condition. Do not use an oxygen coupler with a missing, damaged, or<br />

worn O-ring.<br />